更新分子诊断方法以检测血培养瓶中甲氧西林敏感和耐药金黄色葡萄球菌分离株

Updating Molecular Diagnostics for Detecting Methicillin-Susceptible and Methicillin-Resistant Staphylococcus aureus Isolates in Blood Culture Bottles.
PubMed 文献 · J Clin Microbiol · 2019-10-23

该研究对4株血培养金黄色葡萄球菌及2株对照菌进行全基因组测序,分析Xpert MRSA/SA BC检测出现基因型-表型不符的遗传基础。3株MRSA因SCCmec-orfX连接区插入元件被误判为MSSA,1株因spa缺失出现假阴性;更新后的检测可正确识别这些变异及空 cassette 菌株。在252株MSSA中,3.9%含空SCCmec cassette,1.6%携带SCCM1,<1%有spa缺失,<1%含其他SCCmec变异。

要点

  • SCCmec-orfX连接区插入或spa缺失可导致Xpert MRSA/SA BC假阴性或误判。
  • 更新后的Xpert检测能正确识别spa和SCCmec变异及空cassette菌株。
  • MSSA中干扰Xpert结果的遗传变异总体少见(各变异均<4%)。

核心结论

  • 导致Xpert MRSA/SA BC误判的遗传变异在MSSA中仍少见,更新后的检测可纠正这些误判。

研究设计与样本

  • 对4株血培养金黄色葡萄球菌和2株对照菌进行全基因组测序,分析Xpert MRSA/SA BC检测的基因型-表型差异。
  • 另对来自美国和欧洲的252株MSSA分离株进行变异筛查。

主要发现

  • 3株MRSA分别在SCCmec-orfX连接区有不同遗传元件插入,被误分类为MSSA。
  • 1株因spa缺失导致假S. aureus阴性结果。
  • 含空SCCmec cassette(无mecA)的对照菌株被Xpert正确判为MSSA。
  • 252株MSSA中,3.9%含空SCCmec cassette,1.6%携带SCCM1,<1%有spa缺失,<1%含其他SCCmec变异。
  • 更新后的Xpert MRSA/SA BC检测成功检出所有spa和SCCmec变异的MRSA,并将空cassette菌株正确鉴定为MSSA。

临床怎么用

  • 临床微生物实验室应关注分子检测因靶序列变异导致的假阴性或误分类,必要时使用更新版本检测。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Molecular diagnostic tests can be used to provide rapid identification of staphylococcal species in blood culture bottles to help improve antimicrobial stewardship. However, alterations in the target nucleic acid sequences of the microorganisms or their antimicrobial resistance genes can lead to false-negative results. We determined the whole-genome sequences of 4 blood culture isolates of Staphylococcus aureus and 2 control organisms to understand the genetic basis of genotype-phenotype discrepancies when using the Xpert MRSA/SA BC test (in vitro diagnostic medical device [IVD]). Three methicillin-resistant S. aureus (MRSA) isolates each had a different insertion of a genetic element in the staphylococcal cassette chromosome (SCCmec)-orfX junction region that led to a misclassification as methicillin-susceptible S. aureus (MSSA). One strain contained a deletion in spa, which produced a false S. aureus-negative result. A control strain of S. aureus that harbored an SCCmec element but no mecA (an empty cassette) was correctly called MSSA by the Xpert test. The second control contained an SCCM1 insertion. The updated Xpert MRSA/SA BC test successfully detected both spa and SCCmec variants of MRSA and correctly identified empty-cassette strains of S. aureus as MSSA. Among a sample of 252 MSSA isolates from the United States and Europe, 3.9% contained empty SCCmec cassettes, 1.6% carried SCCM1, <1% had spa deletions, and <1% contained SCCmec variants other than those with SCCM1 These data suggest that genetic variations that may interfere with Xpert MRSA/SA BC test results remain rare. Results for all the isolates were correct when tested with the updated assay.