文献速递临床生化质量管理与危急值

Barricor采血管与PST在多种生化和免疫分析物上等效,但乳酸脱氢酶除外

Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase.
PubMed 文献 · Clin Chim Acta · 2019-06-12

在次优离心条件下,除乳酸脱氢酶(LD)外所有分析物均无偏倚;Barricor管在1912、2060和4000×g离心10分钟时LD较PST显著升高(+7.4-19.4%),且小体积管亦与较高LD相关。作者建议应仔细评估BD离心方案以确定Barricor在本地前处理配置下是否等效于PST。

要点

  • Barricor管在1912、2060和4000×g离心10分钟时,LD较PST升高+7.4-19.4%。
  • 较小管体积(4.5 ml)与Barricor管中LD升高相关,而钾、磷酸盐和AST无变化。

核心结论

  • Barricor管在多种生化和免疫分析物上与PST等效,但乳酸脱氢酶(LD)对离心条件变化最敏感,需评估本地离心方案。

研究设计与样本

  • 从4家本地医院采集配对的PST和Barricor样本,基于站点特定的自动化或手动离心系统处理。
  • 离心条件包括1912×g 10分钟(次优)、2060g 10分钟(替代)和4000×g 3或10分钟(最优),并评估管体积(4.5 vs 5.5 ml)。

主要发现

  • 在自动化前处理系统次优离心条件下,除LD外所有测试分析物在PST和Barricor之间无偏倚。
  • Barricor管在1912、2060和4000×g离心10分钟时,LD较PST显著升高,范围为+7.4-19.4%(整个测量区间87-493 U/l)。
  • 较小管体积也与较高的LD相关,尽管其他溶血标志物如钾、磷酸盐和AST无变化。

临床怎么用

  • 在将Barricor管替代PST前,应基于本地前处理配置(离心速度和时间)仔细评估BD离心方案,尤其关注LD结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: The BD Barricor tube uses a novel mechanical separator designed to eliminate gel artifacts, decrease cellular contamination, and improve stability. Here, we evaluated the Barricor tube as a possible replacement for PST using Beckman Coulter analyzers under both optimal, alternative, and suboptimal centrifugation conditions based on BD recommendations. METHODS: Paired PST and Barricor samples were collected from 4 local hospitals and processed based on site-specific preanalytical systems involving automated or manual centrifugation. Centrifugation conditions ranged from 1912 ×g for 10 min (suboptimal), 2060 g for 10 min (alternative), and 4000 ×g for 3 or 10 min (optimal). Tube volume (4.5 vs. 5.5 ml) was also assessed. Forty-three chemistry and immunochemistry analytes were measured on Beckman Coulter DxC and DxI analyzers. RESULTS: Using an automated preanlaytical system with suboptimal spin conditions, no bias between PST and Barricor was observed for all analytes tested except lactate dehydrogenase (LD). Further investigation revealed significant increase in LD when Barricor was spun for 10 min at 1912, 2060 and 4000 ×g, ranging from +7.4-19.4% vs. PST across the entire measurement interval (87-493 U/l). Smaller tube volume was also associated with higher LD. Differences in LD occurred despite no change in other hemolysis markers such as potassium, phosphate, and AST. CONCLUSIONS: LD is most sensitive to varying centrifugation conditions (time and speed) in Barricor tubes. We recommend that BD centrifugation protocols should be closely evaluated to determine if Barricor is equivalent to PST under local preanalytical configurations.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Raizman JE, Goudreau BL, Füzéry AK, et al. Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase. Clin Chim Acta. 2019. PMID: 31201816. doi:10.1016/j.cca.2019.06.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。