全自动多重梅毒螺旋体与非螺旋体(RPR)检测方法的评估
该研究比较了BioPlex 2200 Syphilis Total & RPR全自动多重流式免疫分析法与LIAISON梅毒螺旋体检测及手工RPR卡片试验的性能,共检测314份血清标本。结果显示BioPlex与LIAISON的总体符合率为96.1%,BioPlex RPR的敏感性和特异性分别为90.5%和97.2%(以手工RPR为金标准)。结论认为BioPlex性能与对照方法相当,且自动化带来劳动节省和流程改进。
要点
- BioPlex全自动梅毒检测与LIAISON螺旋体检测总体符合率96.1%。
- BioPlex RPR敏感性90.5%、特异性97.2%(以手工RPR为金标准)。
- 自动化RPR检测可节省人力、报告客观并改善工作流程。
核心结论
- BioPlex 2200 Syphilis Total & RPR检测性能与LIAISON螺旋体检测及手工RPR试验相当。
研究设计与样本
- 研究纳入314份血清标本,同时用LIAISON螺旋体检测、BioPlex 2200 Syphilis Total & RPR及手工BD Macro-Vue RPR卡片试验检测,不一致结果用T. pallidum颗粒凝集试验进一步确认。
主要发现
- BioPlex检测螺旋体抗体与LIAISON的总体符合率为96.1%。
- 以手工RPR为金标准,BioPlex RPR的敏感性为90.5%,特异性为97.2%。
临床怎么用
- BioPlex全自动梅毒检测可用于同时检测螺旋体和非螺旋体抗体及RPR滴度,自动化流程可减少手工操作并提高报告客观性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: In June 2017, Bio-Rad Laboratories received US Food and Drug Administration clearance for its BioPlex 2200 Syphilis Total & RPR (rapid plasma reagin) assay. It is the first fully automated treponemal/nontreponemal multiplex flow immunoassay, simultaneously detecting Treponema pallidum and reagin antibodies and an RPR titer. We compared the performance of the BioPlex Syphilis Total & RPR assay with the LIAISON Treponema Assay and the manual BD Macro-Vue RPR 18-mm Circle Test. METHODS: In total, 314 serum specimens were tested for treponemal immunoglobulin G/immunoglobulin M and RPR with the LIAISON Treponema Assay, the BioPlex 2200 Syphilis Total & RPR assay, and the manual BD Macro-Vue RPR card test. All discordant results were further tested with the T pallidum particle agglutination assay from Fujirebio Diagnostics. RESULTS: The overall percent agreement for the BioPlex assay for treponemal antibodies with the LIAISON Treponema Assay was 96.1%. Sensitivity and specificity for the BioPlex RPR assay were 90.5% and 97.2%, respectively (the manual RPR assay was considered the gold standard). CONCLUSIONS: The BioPlex 2200 Syphilis Total & RPR assay performance was comparable to the LIAISON Treponema Assay and the manual RPR test. Compared with the manual RPR, the automation of RPR testing offered labor savings, objective result reporting, and improved workflow.
引用
- Arbefeville S, Lynch M, Ferrieri P, et al. Evaluation of a Multiplex Fully Automated Treponemal and Nontreponemal (Rapid Plasma Reagin) Assay. Am J Clin Pathol. 2019. PMID: 31139835. doi:10.1093/ajcp/aqz034
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。