文献速递质量管理与危急值临床生化

参考物质与质控物质的互换性:确保检验医学测量质量的关键因素

Commutability of reference and control materials: an essential factor for assuring the quality of measurements in Laboratory Medicine.
PubMed 文献 · Clin Chem Lab Med · 2019-01-01

溯源至共同参考物质可保证不同检测方法结果的一致性,而溯源依赖互换性良好的参考物质。使用非互换性参考物质校准会引入偏倚,导致临床样本结果错误。CLSI指南要求选取20份覆盖相关浓度范围的临床样本,用两种方法分析参考物质和临床样本,通过回归分析判断参考物质是否落在临床样本定义的95%预测区间内。IFCC工作组近期完善了互换性评估建议,且互换性也是外部质量评估结果可用于评价实验室标准化表现的前提,但EQA材料通常未经验证互换性。

要点

  • 溯源链依赖互换性参考物质,非互换性物质校准会引入偏倚。
  • CLSI互换性评估需20份临床样本、双方法分析和回归分析。
  • EQA材料互换性常未验证,影响其评价实验室标准化表现。

核心结论

  • 参考物质与质控物质的互换性是确保检验医学测量质量的关键因素,非互换性参考物质校准会导致临床样本结果偏倚。

研究设计与样本

  • 综述性文章,阐述互换性概念、CLSI指南评估方法及IFCC工作组近期建议。

主要发现

  • 1973年互换性定义为酶物质显示与血清中同种酶可比的批间活性变化,后扩展为所有参考物质的特性。
  • CLSI指南要求选取20份覆盖相关浓度范围的临床样本,用两种方法分析参考物质和临床样本,通过回归分析判断参考物质是否落在临床样本定义的95%预测区间内。
  • IFCC工作组近期完善了互换性评估建议。
  • EQA材料通常未经验证互换性,影响其用于评价实验室标准化表现。

关键数值与推荐

  • CLSI互换性评估需20份临床样本,覆盖相关浓度范围。
  • 使用回归分析计算参考物质是否落在临床样本定义的95%预测区间内。

临床怎么用

  • 选择互换性良好的参考物质进行校准,避免使用非互换性物质引入偏倚。
  • 评估EQA材料互换性,确保其结果可用于评价实验室标准化表现。

局限性

  • 互换性概念仍被理解不足,CLSI方法受到批评,EQA材料互换性常未验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Traceability to a common reference ensures equivalence of results obtained by different assays. Traceability is achieved by an unbroken sequence of calibrations, using reference materials (RMs) that must be commutable. Using non-commutable RMs for calibration will introduce a bias in the calibrated method producing incorrect results for clinical samples (CS). Commutability was defined in 1973 as "the ability of an enzyme material to show inter-assay activity changes comparable to those of the same enzyme in human serum" and later extended as a characteristic of all RMs. However, the concept is still poorly understood and appreciated. Commutability assessment has been covered in CLSI guidelines and requires: (a) selection of 20 CS spanning the relevant concentration range; (b) analysis of both RM and CS with the pair of procedures; (c) data elaboration using regression analysis and calculation if RM fall within the 95% prediction interval defined by CS. This approach has been criticized and to improve it The International Federation of Clinical Chemistry and Laboratory Medicine established a working group that recently finalized recommendations. Commutability is also a requirement for the applicability of external quality assessment (EQA) results in the evaluation of the performance of participating laboratories in terms of standardization of their measurements. Unfortunately, EQA materials are usually not validated for commutability.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Braga F, Panteghini M. Commutability of reference and control materials: an essential factor for assuring the quality of measurements in Laboratory Medicine. Clin Chem Lab Med. 2019. PMID: 30903757. doi:10.1515/cclm-2019-0154

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。