高敏cTnI方法第99百分位参考上限与参考变化值的多中心评估
意大利SIBioC与ELAS促进的多中心研究纳入1306名健康成人,评估Access高敏肌钙蛋白I(hs-cTnI)方法的第99百分位参考上限(URL)和参考变化值(RCV)。结果显示该方法满足国际指南对高敏方法的两项要求,且第99百分位URL受年龄、性别及统计方法影响显著。
要点
- Access hsTnI方法满足国际指南高敏cTnI方法的两个标准。
- 第99百分位URL受参考人群年龄和性别影响。
- 第99百分位URL亦受计算cTnI分布参数的统计方法影响。
核心结论
- Access hsTnI方法使用DxI平台满足国际指南对高敏cTnI方法的两项要求。
- 第99百分位URL受年龄、性别及统计方法显著影响。
研究设计与样本
- 意大利多中心研究,8个临床中心共收集1306名健康成人志愿者的肝素化血浆样本。
- 每个中心收集50至150份健康成人血浆样本。
- 另有一个457名成人(平均年龄63.0岁,SD 8.1岁,最小47岁,最大86岁)的较老队列接受心电图和心脏影像分析以排除无症状心脏病。
主要发现
- Access hsTnI方法满足国际指南对高敏cTnI方法的两个标准。
- 第99百分位URL计算受参考人群年龄和性别影响。
- 第99百分位URL计算亦受统计方法影响。
关键数值与推荐
- 健康成人志愿者1306名,来自8个意大利临床中心,每个中心收集50至150份样本。
- 较老队列457名成人,平均年龄63.0岁(SD 8.1岁),范围47至86岁。
临床怎么用
- 采用Access hsTnI方法时需考虑年龄和性别对第99百分位URL的影响。
- 计算第99百分位URL时应关注统计方法的选择对结果的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The Italian Society of Clinical Biochemistry (SIBioC) and the Italian Section of the European Ligand Assay Society (ELAS) have recently promoted a multicenter study (Italian hs-cTnI Study) with the aim to accurately evaluate analytical performances and reference values of the most popular cTnI methods commercially available in Italy. The aim of this article is to report the results of the Italian hs-cTnI Study concerning the evaluation of the 99th percentile URL and reference change (RCV) values around the 99th URL of the Access cTnI method. MATERIALS AND METHODS: Heparinized plasma samples were collected from 1306 healthy adult volunteers by 8 Italian clinical centers. Every center collected from 50 to 150 plasma samples from healthy adult subjects. All volunteers denied the presence of chronic or acute diseases and had normal values of routine laboratory tests (including creatinine, electrolytes, glucose and blood counts). An older cohort of 457 adult subjects (mean age 63.0 years; SD 8.1 years, minimum 47 years, maximum 86 years) underwent also ECG and cardiac imaging analysis in order to exclude the presence of asymptomatic cardiac disease. RESULTS AND CONCLUSIONS: The results of the present study confirm that the Access hsTnI method using the DxI platform satisfies the two criteria required by international guidelines for high-sensitivity methods for cTn assay. Furthermore, the results of this study confirm that the calculation of the 99th percentile URL values are greatly affected not only by age and sex of the reference population, but also by the statistical approach used for calculation of cTnI distribution parameters.
引用
- Clerico A, Ripoli A, Masotti S, et al. Evaluation of 99th percentile and reference change values of a high-sensitivity cTnI method: A multicenter study. Clin Chim Acta. 2019. PMID: 30826369. doi:10.1016/j.cca.2019.02.029
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。