文献速递临床微生物质量管理与危急值

半自动药敏判读硬件用于非结核分枝杆菌可能高估敏感性

Implementation of Semiautomated Antimicrobial Susceptibility Interpretation Hardware for Nontuberculous Mycobacteria May Overestimate Susceptibility.
PubMed 文献 · J Clin Microbiol · 2019-03-28

该盲法试验比较手工判读与SensiTitre Vizion半自动判读对132株非结核分枝杆菌(NTM)的药敏试验结果,共727项测试覆盖13种有CLSI折点的抗生素。不一致主要源于CLSI折点接近或重叠MIC50值,在标准临床诊断中使用Vizion可能高估NTM的抗生素敏感性。

要点

  • Vizion半自动判读所得MIC显著低于手工判读(P<0.0001)。
  • 标准临床诊断中使用Vizion可能高估NTM的抗生素敏感性。

核心结论

  • 使用SensiTitre Vizion进行NTM药敏试验可能高估抗生素敏感性。

研究设计与样本

  • 盲法试验比较手工判读与SensiTitre Vizion半自动判读对132株NTM分离株的药敏试验结果,共727项测试覆盖13种有CLSI折点的抗生素。

主要发现

  • Vizion辅助药敏试验产生的MIC显著更低(配对Wilcox符号秩检验;P<0.0001)。
  • 计算得特异度0.8370、灵敏度0.9550、阳性预测值0.8460、阴性预测值0.9520、F1值0.8970。
  • 不一致判读主要源于CLSI指南折点接近或重叠MIC50值,导致跨折点的小差异, contribute to VMEs和MEs。

关键数值与推荐

  • CLSI指南批准的微量滴定板手工判读用于确定抗生素最低抑菌浓度(MIC)。
  • 13种抗生素具有已建立的CLSI折点。

临床怎么用

  • 在NTM标准临床诊断中使用Vizion可能高估抗生素敏感性,需谨慎解读半自动判读结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Nontuberculous mycobacteria (NTM) cause severe opportunistic infections and have a rising incidence in most settings. Rising diagnostic need must be met by national reference laboratories, which rely on Clinical and Laboratory Standards Institute (CLSI) guideline-approved manual readout of microtiter plates for antimicrobial susceptibility testing (AST) to determine antibiotic minimum inhibitory concentrations (MICs). Interpretation of these plates leads to different outcomes between laboratories. The SensiTitre Vizion digital MIC viewing system (Vizion) offers a more streamlined approach using semiautomated reading. Here, we conducted a blinded trial comparing the outcome of AST between manual readout and Vizion readout for 132 NTM isolates, amounting to 727 individual tests for antibiotic susceptibility ranging across 13 individual antibiotics with established CLSI breakpoints. From this, we calculated specificity, sensitivity, positive predictive value (PPV), negative predictive value (NPV) and the F1 value, as well as assessing major error (ME) and very major error (VME) rates. We find that Vizion-assisted AST produces significantly lower MICs (paired Wilcox signed rank test; P < 0.0001). The Vizion had an accuracy of 89,40%, producing 61 MEs (8.39%) and 16 VMEs (2.20%). The calculated specificity was 0.8370, the sensitivity was 0.9550, the PPV was 0.8460, the NPV was 0.9520, and the F1 score was 0.8970. We show that discrepant readings mostly stem from CLSI guideline breakpoints being close to, or overlapping, the MIC50 values, leading to small discrepancies crossing the breakpoint, contributing to VMEs and MEs. Using the Vizion in standard clinical diagnostics for NTM might lead to an overestimation of antibiotic susceptibility.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Rockland M, Ruth MM, Aalders N, et al. Implementation of Semiautomated Antimicrobial Susceptibility Interpretation Hardware for Nontuberculous Mycobacteria May Overestimate Susceptibility. J Clin Microbiol. 2019. PMID: 30760530. doi:10.1128/JCM.01756-18

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。