六种自动化免疫分析法与ID-LC-MS/MS测定总甲状腺素的比较

Comparison of Six Automated Immunoassays With Isotope-Diluted Liquid Chromatography-Tandem Mass Spectrometry for Total Thyroxine Measurement.
PubMed 文献 · Ann Lab Med · 2019-01-01

该研究以ID-LC-MS/MS为参考方法,比较六种自动化免疫分析法测定156例患者血清总甲状腺素(TT4)的表现,并评估TSH与TT4的相关性。结果显示所有免疫分析法与参考方法相关性>0.945,但多数不符合基于生物学变异的最低临床要求(平均偏倚≤4.5%、总CV≤3.7%)。

要点

  • 所有免疫分析法与ID-LC-MS/MS相关性>0.945,但斜率和截距差异明显。
  • 仅Autobio、Beckman-Coulter和Roche的斜率95%CI包含1。
  • 仅Abbott和Roche的批内和总CV≤3.7%。

核心结论

  • 所有免疫分析法与ID-LC-MS/MS强相关,但多数未满足最低临床要求。
  • 实验室和免疫分析制造商需认识到这些局限性。

研究设计与样本

  • 收集2015年10月至2016年1月156例患者血清。
  • 用六种自动化免疫分析法和ID-LC-MS/MS测定TT4,采用Passing-Bablok回归和Bland-Altman分析。
  • 基于生物学变异的最低要求:平均偏倚≤4.5%,总CV≤3.7%。

主要发现

  • 所有免疫分析法与ID-LC-MS/MS相关性>0.945。
  • Passing-Bablok回归斜率从0.886(Mindray)到1.23(Siemens),截距从-12.8(Siemens)到4.61(Mindray)。
  • 仅Autobio、Beckman-Coulter和Roche的斜率95%CI包含1。
  • 平均偏倚从-10.8%(Abbott)到9.0%(Siemens),Roche最低为3.5%,Abbott最高为-10.8%。
  • 仅Abbott和Roche的批内和总CV≤3.7%。

关键数值与推荐

  • 最低要求:平均偏倚≤4.5%,总CV≤3.7%。

临床怎么用

  • 实验室应关注所用TT4免疫分析法的偏倚和精密度是否满足临床要求。
  • 免疫分析制造商需改进方法性能以符合最低临床要求。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Accurate serum total thyroxine (TT4) measurement is important for thyroid disorder diagnosis and management. We compared the performance of six automated immunoassays with that of isotope-diluted liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) as the reference method. We also evaluated the correlation of thyroid stimulating hormone (TSH) with TT4 measured by ID-LC-MS/MS and immunoassays. METHODS: Serum was collected from 156 patients between October 2015 and January 2016. TT4 was measured by immunoassays from Abbott (Architect), Siemens (ADVIA Centaur XP), Roche (E601), Beckman-Coulter (Dxi800), Autobio (Autolumo A2000), and Mindray (CL-1000i), and by ID-LC-MS/MS. Results were analyzed using Passing-Bablok regression and Bland-Altman plots. Minimum requirements based on biological variation were as follows: a mean bias of ≤4.5% and total imprecision (CV) of ≤3.7%. RESULTS: All immunoassays showed a correlation >0.945 with ID-LC-MS/MS; however, the slope of the Passing-Bablok regression line varied from 0.886 (Mindray) to 1.23 (Siemens) and the intercept from -12.8 (Siemens) to 4.61 (Mindray). Only Autobio, Beckman-Coulter, and Roche included the value of one in the 95% confidence interval for slope. The mean bias ranged from -10.8% (Abbott) to 9.0% (Siemens), with the lowest value noted for Roche (3.5%) and the highest for Abbott (-10.8%). Only Abbott and Roche showed within-run and total CV ≤3.7%. CONCLUSIONS: Though all immunoassays correlated strongly with ID-LC-MS/MS, most did not meet the minimum clinical requirement. Laboratories and immunoassay manufacturers must be aware of these limitations.