文献速递临床免疫质量管理与危急值

抗PR3 ANCA IgG自身抗体的有证参考物质:从开发到认证

Certified reference material against PR3 ANCA IgG autoantibodies. From development to certification.
PubMed 文献 · Clin Chem Lab Med · 2019-01-01

该研究评估了10种免疫测定法检测30份临床样本中PR3 ANCA IgG的结果差异,发现6份样本结果分类不一致。基于此开发并认证了一种适用于校准的互通性有证参考物质,以期改善不同测定法结果的可比性。

要点

  • 不同PR3 ANCA IgG免疫测定法对部分临床样本结果分类不一致。
  • 8种测定法组合间临床样本结果相关性系统良好。
  • 使用互通性有证参考物质校准可改善结果可比性。

核心结论

  • 使用互通性有证参考物质校准应能改善PR3 ANCA IgG免疫测定法结果的可比性。

研究设计与样本

  • 用10种免疫测定法评估30份临床样本,并两两评估测定法间结果相关性。
  • 开展可行性研究以选择最适合制备有证参考物质的参考物质形式。

主要发现

  • 两套各30份临床样本用10种测定法评估显示,测定法间差异可导致个别临床样本解释不同。
  • 多数样本在所有测定法中结果分类相同,但6份样本检测导致结果不一致。
  • 8种测定法组合间临床样本结果相关性系统良好。

临床怎么用

  • 检验科智能体医生应关注PR3 ANCA IgG不同免疫测定法间结果差异对临床解释的影响。
  • 采用有证参考物质校准有助于提升不同测定法结果的可比性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Background The importance of the standardisation of immunoassays for autoantibodies has been widely discussed. The appropriate use of certified reference materials (CRM) could contribute to a more accurate diagnosis and follow-up of a series of diseases such as small vessel-associated vasculitis. This is a systemic autoimmune disorder during which two autoantibodies can be present, MPO ANCA IgG and PR3 ANCA IgG. Results from different commercially available immunoassays used for PR3 ANCA IgG measurement can vary significantly. Therefore the potential for improvement using a suitable certified reference material was assessed and led to the development of a CRM. Methods Thirty clinical samples were evaluated using 10 immunoassays. The correlation between results from these assays was assessed in a pairwise manner. Feasibility studies were conducted in order to find a reference material format most suitable for the preparation of a CRM. Results The evaluation of two sets of 30 clinical samples with 10 assays showed that differences between assays can result in different interpretations for individual clinical samples. Most of the samples had the same result classification in all assays. However, six of the samples tested led to inconsistent results. Conclusions The correlation between results from clinical samples was systematically good for combinations of eight of those assays. Therefore, it should be possible to improve the comparability of results using a commutable CRM for calibration. Based on these studies, a final format for the CRM was selected and eventually produced and certified for its PR3 ANCA IgG content.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Monogioudi E, Sheldon J, Meroni PL, et al. Certified reference material against PR3 ANCA IgG autoantibodies. From development to certification. Clin Chem Lab Med. 2019. PMID: 30789822. doi:10.1515/cclm-2018-1095

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。