BacT/Alert Virtuo与BacT/Alert 3D血培养系统临床性能比较
该研究纳入952例患者共1904套血培养,比较BacT/Alert Virtuo与3D系统的阳性率和检出时间。两系统阳性率无显著差异,但Virtuo对所有分离菌、金黄色葡萄球菌和大肠埃希菌的中位检出时间均更短。
要点
- Virtuo与3D系统阳性率无显著差异。
- Virtuo对所有分离菌中位检出时间短于3D系统。
- Virtuo对金黄色葡萄球菌和大肠埃希菌检出更快。
核心结论
- Virtuo对所有瓶型中的病原体均具有早期检出潜力,可能通过快速管理改善脓毒症预后。
研究设计与样本
- 纳入韩国大学附属医院952例患者共1904套血培养,每套向Virtuo和3D系统的需氧瓶(FA Plus)和厌氧瓶(FN Plus)中各注入5 mL血液,比较阳性率和检出时间。
主要发现
- FA Plus瓶阳性率在Virtuo和3D系统分别为8.3%和8.4%,FN Plus瓶分别为7.8%和8.3%(P>0.05)。
- 所有分离菌中位检出时间Virtuo为11.5小时(N=305),3D为11.8小时(N=318),P<0.001。
- 金黄色葡萄球菌中位检出时间Virtuo为14.3小时,3D为16.0小时(N=38,P=0.021)。
- 大肠埃希菌中位检出时间Virtuo为10.4小时,3D为11.0小时(N=117,P<0.001)。
临床怎么用
- Virtuo系统可更早检出病原体,有助于脓毒症患者快速启动管理。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: BacT/Alert Virtuo (BioMérieux, Durham, NC, USA) is a recently developed blood culture system that includes functions of automatic registration, loading, and unloading of the blood culture bottles, as well as measurement of blood volume. We compared the performances between the BacT/Alert Virtuo and 3D (BioMérieux) blood culture systems. METHODS: A total of 952 patients (1,904 sets) visiting an university-affiliated hospital in Korea for blood cultures were enrolled. Five milliliters of blood was added into each of the two aerobic (FA Plus) and two anaerobic (FN Plus) bottles of the Virtuo and 3D systems for a single set. Positive rate and time to detection (TTD) were compared between the two systems. RESULTS: The positive rates were 8.3% and 8.4% in FA Plus bottles and 7.8% and 8.3% in FN Plus bottles, in the Virtuo and 3D systems, respectively (P>0.05). Median TTDs were shorter in the Virtuo than in the 3D system for all isolates (11.5 hours [N=305] vs 11.8 hours [N=318], P<0.001), Staphylococcus aureus (N=38; 14.3 hours vs 16.0 hours, P=0.021), and Escherichia coli (N=117; 10.4 hours vs 11.0 hours, P<0.001). CONCLUSIONS: The Virtuo has the potential to detect pathogens early in all bottle types. This might improve the prognosis of sepsis by allowing for implementation of expeditious management.
引用
- Kim SC, Lee S, Kim S, et al. Comparison of Clinical Performance Between BacT/Alert Virtuo and BacT/Alert 3D Blood Culture Systems. Ann Lab Med. 2019. PMID: 30623614. doi:10.3343/alm.2019.39.3.278