文献速递临床生化临床免疫

韩国新生儿干血斑LC-MS/MS多类固醇谱分析

Dried Blood Spot Multiplexed Steroid Profiling Using Liquid Chromatography Tandem Mass Spectrometry in Korean Neonates.
PubMed 文献 · Ann Lab Med · 2019-01-01

该研究开发并验证了一种液相色谱-串联质谱(LC-MS/MS)方法,可同时定量干血斑(DBS)中九种类固醇激素,并在1146例韩国新生儿(272例早产儿和874例足月儿)中建立了参考区间。与免疫分析法相比,LC-MS/MS检测17α-羟孕酮存在负偏倚(免疫法测值更高,r=0.8104,P<0.0001),并成功应用于14例额外样本的先天性肾上腺皮质增生症(CAH)筛查与诊断。

要点

  • 建立了DBS中九种类固醇的LC-MS/MS定量方法并验证了性能。
  • 提供了韩国早产儿和足月儿九种类固醇的参考区间。
  • 免疫法测17α-羟孕酮较LC-MS/MS偏高,提示假阳性风险。

核心结论

  • LC-MS/MS方法可高灵敏、高特异地从DBS中评估九种类固醇,是CAH临床分析的有前景工具。

研究设计与样本

  • 研究开发并验证了LC-MS/MS方法,评估线性、精密度、准确性、提取回收率和基质效应,并在1146例DBS样本(272例早产儿和874例足月儿)中确定参考区间。
  • 对17α-羟孕酮比较了免疫分析法与LC-MS/MS法,并用14例额外样本验证临床适用性。

主要发现

  • 线性范围:皮质醇2.8-828.0 nmol/L,其他类固醇0.9-40.0 nmol/L(R²>0.99)。
  • 日内和日间精密度CV分别为2.52-12.26%和3.53-17.12%。
  • 准确性为80.81-99.94%,提取回收率为88.0-125.4%,基质效应为61.7-74.2%。
  • 免疫法测值较LC-MS/MS存在负偏倚(免疫法更高,r=0.8104,P<0.0001)。
  • LC-MS/MS方法成功应用于14例额外样本的九种类固醇分析,用于CAH筛查和诊断。

关键数值与推荐

  • 皮质醇线性范围2.8-828.0 nmol/L,其他类固醇0.9-40.0 nmol/L(R²>0.99)。
  • 日内精密度CV 2.52-12.26%,日间精密度CV 3.53-17.12%。
  • 准确性80.81-99.94%,提取回收率88.0-125.4%,基质效应61.7-74.2%。
  • 免疫法与LC-MS/MS测17α-羟孕酮相关性r=0.8104,P<0.0001。

临床怎么用

  • LC-MS/MS法可用于新生儿DBS中九种类固醇的同时定量,辅助CAH筛查与诊断。
  • 免疫法测17α-羟孕酮存在高偏倚,LC-MS/MS可减少假阳性。
  • 参考区间基于韩国早产儿和足月儿DBS样本建立,临床解读时需考虑人群和胎龄。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Screening for congenital adrenal hyperplasia (CAH) using immunoassays for 17α-hydroxyprogesterone generates many false-positive results. We developed and validated a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for simultaneous quantification of nine steroid hormones in dried blood spot (DBS) samples, and established reference intervals for these hormones. METHODS: We examined our method for linearity, precision, accuracy, extraction recovery, and matrix effects and determined the reference intervals of cortisol, 17α-hydroxyproges-terone, 11-deoxycortisol, 21-deoxycortisol, androstenedione, corticosterone, 11-deoxycorticosterone, testosterone, and progesterone in 1,146 DBS samples (from 272 preterm and 874 full-term neonates). Immunoassay and LC-MS/MS methods were compared for 17α-hydroxyprogesterone. Fourteen additional samples were tested to validate the clinical applicability of the LC-MS/MS method. RESULTS: The linearity range was 2.8-828.0 nmol/L for cortisol and 0.9-40.0 nmol/L for the other steroids (R²>0.99). Intra-day and inter-day precision CVs were 2.52-12.26% and 3.53-17.12%, respectively. Accuracy was 80.81-99.94%, and extraction recovery and matrix effects were 88.0-125.4% and 61.7-74.2%, respectively. There was a negative bias, with higher values measured by immunoassay compared with LC-MS/MS (r=0.8104, P<0.0001). The LC-MS/MS method was successfully applied to the analysis of nine steroids in DBS for screening and diagnosis of CAH using the 14 additional samples. CONCLUSIONS: Our method enables highly sensitive and specific assessment of nine steroids from DBS and is a promising tool for clinical analysis of CAH.

来源:Ann Lab Med,查看原文。版权归原作者所有。

引用

  1. Choi R, Park HD, Oh HJ, et al. Dried Blood Spot Multiplexed Steroid Profiling Using Liquid Chromatography Tandem Mass Spectrometry in Korean Neonates. Ann Lab Med. 2019. PMID: 30623618. doi:10.3343/alm.2019.39.3.263

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。