CALIPER健康儿童青少年Siemens ADVIA XPT/1800与Dimension EXL生化检测儿童参考区间建立
本研究基于CALIPER健康儿童青少年队列,在Siemens ADVIA XPT/1800和Dimension EXL系统上分别检测909例和867例0-<19岁血清样本,为54项生化检测建立年龄和/或性别特异性参考区间。碱性磷酸酶和肌酐等显示年龄与性别差异,免疫球蛋白A和铁仅显示年龄差异,钠和甘油三酯在儿童年龄范围内相对恒定且性别间相似。
要点
- 建立Siemens ADVIA XPT/1800和Dimension EXL系统54项生化检测的儿童年龄和/或性别特异性参考区间。
- 碱性磷酸酶和肌酐显示年龄和性别差异,免疫球蛋白A和铁仅显示年龄差异。
- 首次利用CALIPER队列建立肌酸激酶、随机血糖、总铁结合力和多种电解质的参考区间。
核心结论
- 本研究建立的儿童参考区间将有助于全球使用Siemens生化检测时更准确地解读实验室检测结果。
研究设计与样本
- 基于CALIPER健康儿童青少年队列,ADVIA XPT/1800系统检测909例0-<19岁血清样本,Dimension EXL系统检测867例0-<19岁血清样本。
主要发现
- 钠和甘油三酯等血清浓度在儿童年龄范围内相对恒定且性别间相似。
- 碱性磷酸酶和肌酐显示年龄和性别差异。
- 免疫球蛋白A和铁仅显示年龄差异。
关键数值与推荐
- 为54项生化检测计算年龄和/或性别特异性参考区间。
- 首次利用CALIPER队列建立肌酸激酶、随机血糖、总铁结合力和多种电解质的参考区间。
临床怎么用
- 临床解读Siemens ADVIA XPT/1800和Dimension EXL系统儿童生化检测结果时应使用年龄和/或性别特异性参考区间。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Accurate reference intervals (RIs) are essential for clinical interpretation of laboratory test results; however, major gaps exist in pediatric RIs. The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) has established age- and sex-specific pediatric RIs on various analytical platforms. The current study expands the CALIPER database by establishing age- and sex-specific RIs for biochemical assays on Siemens ADVIA XPT/1800 and Dimension EXL Systems. METHODS: Serum samples from a total of 909 and 867 healthy children and adolescents (ages 0-<19 y) were tested on ADVIA XPT/1800 and Dimension EXL systems, respectively. Age- and/or sex-specific RIs were calculated for a total of 54 biochemical assays. RESULTS: Serum concentrations of several biomarkers remained relatively constant across the pediatric age range and similar between sexes, including sodium and triglycerides. Other biomarkers, such as alkaline phosphatase and creatinine showed both age and sex differences. Furthermore, immunoglobulin A and iron showed only age differences. DISCUSSION: We established RIs for creatine kinase, random glucose, total iron binding capacity, and several electrolytes for the first time using the CALIPER cohort. Overall, pediatric RIs established in the current study will allow for more accurate laboratory test interpretation worldwide using Siemens chemistry assays.
引用
- Tahmasebi H, Higgins V, Woroch A, et al. Pediatric reference intervals for clinical chemistry assays on Siemens ADVIA XPT/1800 and Dimension EXL in the CALIPER cohort of healthy children and adolescents. Clin Chim Acta. 2018. PMID: 30550936. doi:10.1016/j.cca.2018.12.011
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。