文献速递临床免疫质量管理与危急值

生物素对免疫测定干扰的综合评估

Comprehensive assessment of biotin interference in immunoassays.
PubMed 文献 · Clin Chim Acta · 2018-10-05

该研究用混合患者血清配制不同浓度生物素溶液,在Roche Cobas 8000 e602平台上以夹心法或竞争法测定多种分析物,发现夹心法结果被不同程度假性降低,竞争法结果假性升高,其中高敏肌钙蛋白T、促甲状腺激素和促卵泡激素降低最明显,三碘甲状腺原氨酸和维生素D升高最明显。

要点

  • 生物素可干扰夹心法免疫测定导致结果假性降低。
  • 生物素可干扰竞争法免疫测定导致结果假性升高。
  • 对服用生物素患者应谨慎解读异常免疫测定结果。

核心结论

  • 生物素可干扰免疫测定并导致检验结果异常,临床医生解读服用生物素患者的异常结果时应谨慎。

研究设计与样本

  • 使用混合患者血清样本制备不同浓度的生物素溶液,并在Roche Cobas 8000 e602平台上以夹心法或竞争法测定所有分析物。

主要发现

  • 部分夹心法免疫测定结果被不同生物素水平不同程度地假性降低。
  • 部分竞争法免疫测定结果被生物素假性升高。
  • 高敏肌钙蛋白T、促甲状腺激素和促卵泡激素结果假性降低最明显。
  • 三碘甲状腺原氨酸和维生素D结果假性升高最明显。
  • 其他免疫测定结果也受到生物素一定程度的干扰。

临床怎么用

  • 临床医生解读服用生物素患者的异常免疫测定结果时应谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Biotinylated antibodies and analogues are currently used in many immunoassays while biotin is widely used as a dietary supplement. Thus, biotin interference is an emerging issue for clinical laboratories. METHODS: Various concentrations of biotin solutions were prepared using pooled patient serum samples. All analytes were measured by sandwich or competitive immunoassay on the Roche Cobas 8000 e602 platform. RESULTS: Some of the sandwich immunoassay results were falsely decreased to different extents by different biotin levels, while some of the competitive immunoassay results were falsely increased. The most notable false reductions were in high-sensitivity troponin T, thyroid-stimulating hormone, and follicle-stimulating hormone results, while the most notable false increases were in triiodothyronine and vitamin D results. Other immunoassay results were also affected to some extent by biotin interference. CONCLUSIONS: Biotin can interfere in immunoassays and result in aberrant test results. Clinicians should use caution in interpreting abnormal results in patients who ingest biotin.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Li J, Wagar EA, Meng QH, et al. Comprehensive assessment of biotin interference in immunoassays. Clin Chim Acta. 2018. PMID: 30296442. doi:10.1016/j.cca.2018.10.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。