Vitek MS与Bruker MALDI-TOF MS鉴定临床标本中金黄杆菌属及罕见感染报告的比较
该研究比较Vitek MS与Bruker Biotyper对2005至2016年140株临床金黄杆菌分离株的鉴定能力,以16S rRNA基因测序为参考。Bruker Biotyper种水平总体准确率90.7%,Vitek MS为60.5%;两者均未正确鉴定罕见金黄杆菌种。建议更新校正数据库。
要点
- Bruker Biotyper种水平总体准确率90.7%,Vitek MS为60.5%。
- Vitek MS仅正确鉴定2.6%的C. gleum,Bruker为100%。
- 两者均未成功鉴定12株罕见金黄杆菌种。
核心结论
- Bruker Biotyper鉴定金黄杆菌种水平总体准确率高于Vitek MS。
- 两种MALDI-TOF MS系统均无法正确鉴定罕见金黄杆菌种。
- 建议升级校正Vitek MS和Bruker Biotyper数据库以正确鉴定金黄杆菌属。
研究设计与样本
- 检索2005至2016年某医院微生物数据库,纳入140株金黄杆菌临床分离株。
- 以16S rRNA基因测序为参考方法,比较Vitek MS与Bruker Biotyper的鉴定结果。
主要发现
- 16S rRNA测序鉴定78株为C. indologenes,39株为C. gleum,12株为罕见金黄杆菌种,11株仅鉴定至属水平。
- Vitek MS和Bruker Biotyper分别正确鉴定98.7%和100%的C. indologenes分离株。
- Bruker Biotyper正确鉴定100%的C. gleum分离株,Vitek MS仅正确鉴定2.6%。
- 两种系统均未成功鉴定任何罕见金黄杆菌种。
- Vitek MS和Bruker Biotyper种水平总体准确率分别为60.5%和90.7%。
关键数值与推荐
- Vitek MS对C. indologenes鉴定准确率98.7%,Bruker Biotyper为100%。
- Bruker Biotyper对C. gleum鉴定准确率100%,Vitek MS为2.6%。
- Vitek MS种水平总体准确率60.5%,Bruker Biotyper为90.7%。
临床怎么用
- 临床微生物实验室可优先选用Bruker Biotyper鉴定金黄杆菌属,尤其C. gleum。
- 对罕见金黄杆菌种,MALDI-TOF MS结果不可靠,需16S rRNA基因测序确认。
- 应定期升级校正MALDI-TOF MS数据库以提高金黄杆菌鉴定准确性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Matrix-assisted laser desorption ionization-time of flight mass spectrometry is becoming more popular and is replacing traditional identification methods in the clinical microbiology laboratory. We aimed to compare the Vitek mass spectrometry (MS) and Bruker Biotyper systems for the identification of Chryseobacterium isolated from clinical specimens and to report uncommon Chryseobacterium infections in humans. The microbial database from a hospital was searched for records between 2005 and 2016 to identify cultures that yielded Chryseobacterium Species identification by the Vitek MS and Bruker Biotyper systems was compared to identification by 16S rRNA gene sequencing. Over the study period, 140 Chryseobacterium isolates were included. Based on 16S rRNA gene sequencing, 78 isolates were C. indologenes, 39 were C. gleum, 12 were uncommon Chryseobacterium species (C. arthrosphaerae, C. culicis, C. cucumeris, C. bernardetii, C. artocarpi, and C. daecheongense), and the remaining 11 isolates were only identified at the genus level. The Vitek MS and Bruker Biotyper systems correctly identified 98.7% and 100% of C. indologenes isolates, respectively. While the Bruker Biotyper accurately identified 100% of C. gleum isolates, the Vitek MS system correctly identified only 2.6% of isolates from this species. None of the uncommon Chryseobacterium species were successfully identified by either of these two systems. The overall accuracies of Chryseobacterium identification at the species level by the Vitek MS and Bruker Biotyper systems were 60.5% and 90.7%, respectively. An upgrade and correction of the Vitek MS and Bruker Biotyper databases is recommended to correctly identify Chryseobacterium species.
引用
- Lin JN, Teng SH, Lai CH, et al. Comparison of the Vitek MS and Bruker Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Systems for Identification of Chryseobacterium Isolates from Clinical Specimens and Report of Uncommon Chryseobacterium Infections in Humans. J Clin Microbiol. 2018. PMID: 30135228. doi:10.1128/JCM.00712-18
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。