该研究纳入254份单微生物阳性血培养样本,经血琼脂平板短期孵育6小时后,用MicroFlex LT和Vitek-MS进行鉴定,再用Vitek2进行药敏试验。MicroFlex LT和Vitek-MS的正确种水平鉴定率分别为82.3%和78.3%;快速药敏法与常规法总体一致率为97.8%,极重大误差率0.5%、重大误差率0.7%、次要误差率1.0%。该方法可在血培养阳性当天提供鉴定结果,药敏结果比常规法提前一天。

要点

  • 阳性血培养肉汤经血琼脂短期孵育6小时可实现同日鉴定。
  • MicroFlex LT种水平正确鉴定率为82.3%,Vitek-MS为78.3%。
  • 快速药敏法与常规法总体一致率97.8%,极重大误差率0.5%。

核心结论

  • 短期固体培养联合快速鉴定与药敏系统可在血培养阳性当天提供鉴定结果,药敏结果比常规法提前一天。

研究设计与样本

  • 研究纳入254份单微生物阳性血培养样本,在血琼脂平板上孵育6小时后用MicroFlex LT和Vitek-MS鉴定,再用Vitek2进行药敏试验,并与常规方法比较。

主要发现

  • MicroFlex LT的正确种水平鉴定率为82.3%(209/254),Vitek-MS为78.3%(199/254)。
  • 对1,174个微生物/抗菌药物组合,快速药敏法与常规法总体一致率为97.8%(1,148/1,174)。
  • 快速药敏法的极重大误差率为0.5%,重大误差率为0.7%,次要误差率为1.0%。

关键数值与推荐

  • 正确种水平鉴定率:MicroFlex LT 82.3%(209/254),Vitek-MS 78.3%(199/254)。
  • 快速药敏法总体一致率97.8%(1,148/1,174),极重大误差率0.5%,重大误差率0.7%,次要误差率1.0%。

临床怎么用

  • 常规开展该短期孵育方法可在血培养阳性当天提供鉴定结果,药敏结果比常规法提前一天,有助于早期适当抗生素治疗。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Early and appropriate antibiotic treatment improves the clinical outcome of patients with septicemia; therefore, reducing the turn-around time for identification (ID) and antimicrobial susceptibility test (AST) results is essential. We established a method for rapid ID and AST using short-term incubation of positive blood culture broth samples on solid media, and evaluated its performance relative to that of the conventional method using two rapid ID systems and a rapid AST method. METHODS: A total of 254 mono-microbial samples were included. Positive blood culture samples were incubated on blood agar plates for six hours and identified by the MicroFlex LT (Bruker Daltonics) and Vitek-MS (bioMeriéux) systems, followed by AST using the Vitek2 System (bioMeriéux). RESULTS: The correct species-level ID rates were 82.3% (209/254) and 78.3% (199/254) for the MicroFlex LT and Vitek-MS platforms, respectively. For the 1,174 microorganism/antimicrobial agent combinations tested, the rapid AST method showed total concordance of 97.8% (1,148/1,174) with the conventional method, with a very major error rate of 0.5%, major error rate of 0.7%, and minor error rate of 1.0%. CONCLUSIONS: Routine implementation of this short-term incubation method could provide ID results on the day of blood culture-positivity detection and one day earlier than the conventional AST method. This simple method will be very useful for rapid ID and AST of bacteria from positive blood culture bottles in routine clinical practice.

来源:Ann Lab Med,查看原文。版权归原作者所有。

引用

  1. Ha J, Hong SK, Han GH, et al. Same-Day Identification and Antimicrobial Susceptibility Testing of Bacteria in Positive Blood Culture Broths Using Short-Term Incubation on Solid Medium with the MicroFlex LT, Vitek-MS, and Vitek2 Systems. Ann Lab Med. 2018. PMID: 29401558. doi:10.3343/alm.2018.38.3.235