Hevylite 检测法定量 M 蛋白的分析验证
该研究在 166 份常规临床样本和 27 份荷兰外部质量评估(EQA)样本中验证 Hevylite 重链/轻链(HLC)免疫分析的分析性能,并与免疫固定电泳(IFE)及常规 Ig 诊断比较。结果显示 HLC 定量多克隆和单克隆 IgG、IgA、IgM 与常规诊断相关性好,与 IFE 总体一致性达 93%,但 IFE 检测单克隆的灵敏度更高。结论认为 HLC 可准确定量电泳法无法定量的 β 区 M 蛋白。
要点
- HLC 法可准确定量多克隆和单克隆 IgG、IgA、IgM,与常规 Ig 诊断相关性好。
- HLC 比值与 IFE 总体一致性 93%,Cohen κ 系数 0.84。
- HLC 法适用于电泳技术无法定量的 β 区 M 蛋白监测。
核心结论
- HLC 检测法是准确定量 M 蛋白的方法,可改善电泳技术无法定量的 β 区 M 蛋白监测。
研究设计与样本
- 研究检测了 166 份常规临床样本和 27 份来自荷兰 M 蛋白诊断 EQA 的样本(74 家参与实验室)。
- 通过验证 HLC 对之和等于总 Ig 浓度评估分析准确性,并与 IFE 直接方法比较确定 HLC 检测灵敏度。
主要发现
- 27 份 EQA 样本中 HLC 与常规 Ig 诊断比较,多克隆和单克隆 IgG、IgA、IgM 定量 Pearson 相关系数分别为 0.94、0.99、0.99,斜率分别为 0.94、1.07、0.98。
- IFE 与 HLC 比值总体一致性高(93%),Cohen κ 系数为 0.84。
- 两种检测方法间的差异主要源于 IFE 检测单克隆的灵敏度更高。
临床怎么用
- HLC 检测可用于定量 M 蛋白,尤其适用于电泳法无法定量的 β 区 M 蛋白监测。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The heavy/light chain (HLC) immunoassay quantifies the different heavy chain/light chain combinations of each immunoglobulin (Ig) class. This makes the HLC assay suited to quantify monoclonal immunoglobulins (M-protein) and for monitoring of patients with monoclonal gammopathies. This method is particularly advantageous for those samples in which electrophoretic quantification of the M-protein is not possible. METHODS: In this study we tested the analytical performance of the HLC assay in 166 routine clinical samples and in 27 samples derived from the Dutch external quality assessment (EQA) for M-protein diagnostics (74 participating laboratories). Analytical accuracy was assessed by verification that the sum of the HLC-pairs equaled total Ig concentration. Sensitivity of the HLC assay was determined in a direct method comparison with immunofixation electrophoresis (IFE). RESULTS: Comparison of HLC data with routine Ig diagnostics in 27 EQA samples showed very good correlation for both the quantification of polyclonal and monoclonal IgG, IgA and IgM (Pearson correlations [r] were 0.94, 0.99 and 0.99, respectively; slopes were 0.94, 1.07 and 0.98, respectively). The overall concordance between IFE and the HLC ratio was high (93%) with a Cohen κ coefficient of 0.84. Discrepancies between both assays were mainly caused by the higher sensitivity of IFE to detect monoclonality. CONCLUSIONS: We conclude that the HLC assay is an accurate method to quantify M-proteins that can improve monitoring of M-proteins in the beta fraction that cannot be quantified using electrophoretic techniques.
引用
- Jacobs JFM, Haagen IA, Lodder A, et al. Analytical validation of the Hevylite assays for M-protein quantification. Clin Chem Lab Med. 2018. PMID: 29397379. doi:10.1515/cclm-2017-0817
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。