直接血培养纸片扩散法检测革兰阴性菌药敏的初步报告
该研究评估了以阳性血培养肉汤直接接种进行纸片扩散法(dDD)检测革兰阴性菌药敏的性能,采用20株挑战菌株接种至三种血培养系统。去除随机误差后,dDD的分类一致率在BacT/Alert、Bactec和VersaTREK系统分别为87.8%、88.4%和92.2%,无极严重错误,但6小时孵育导致部分抑菌圈无法判读。
要点
- dDD在三种血培养系统中的分类一致率均较高,无极严重错误。
- 6小时孵育导致19.9%的测试因生长过淡无法判读抑菌圈。
- 细菌浓度在7.6×10^7至5.0×10^8 CFU/ml时dDD性能最佳。
核心结论
- 直接血培养纸片扩散法(dDD)在临床实验室中具有潜在准确性。
研究设计与样本
- 研究纳入20株挑战菌株(肠杆菌科、鲍曼不动杆菌和铜绿假单胞菌),接种至人血后注入三种血培养瓶并孵育。
主要发现
- 去除随机误差后,dDD分类一致率在BacT/Alert、Bactec和VersaTREK系统分别为87.8%、88.4%和92.2%。
- 未观察到极严重错误(VME),主要错误(ME)率分别为3.0%、2.3%和1.7%。
- 6小时孵育相比16-18小时,19.9%的测试因生长过淡无法判读抑菌圈。
- 在可评估的dDD测试中,6小时判读时BacT/Alert、Bactec和VersaTREK系统的分类一致率分别为58.9%、76.6%和73.2%。
- 细菌浓度在7.6×10^7至5.0×10^8 CFU/ml时dDD性能最佳,18小时孵育分类一致率为94.7-96.2%,6小时为76.9-84.1%。
关键数值与推荐
- dDD分类一致率:BacT/Alert 87.8%,Bactec 88.4%,VersaTREK 92.2%(去除随机误差后)。
- 极严重错误率:0%;主要错误率:BacT/Alert 3.0%,Bactec 2.3%,VersaTREK 1.7%。
- 6小时判读时,可评估测试的分类一致率:58.9%、76.6%、73.2%;极严重错误率:2.2%、1.8%、3.0%;主要错误率:25.4%、6.1%、2.8%。
临床怎么用
- dDD可作为临床实验室快速药敏检测的潜在方法,但需注意孵育时间和细菌浓度对结果判读的影响。
局限性
- 本研究为初步报告,样本量有限(20株挑战菌株),需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The performance of a disk diffusion test using broth from positive blood cultures as inoculum (direct disk diffusion [dDD]) was evaluated for a collection of 20 challenge isolates of Enterobacteriaceae, Acinetobacter baumannii, and Pseudomonas aeruginosa Isolates seeded into human blood were inoculated into Bactec Plus Aerobic/F, VersaTREK Redox 1, and BacT/Alert FA Plus bottles and incubated in the respective automated blood culture systems. Disk diffusion results were compared to reference disk diffusion results. Categorical agreement (CA) values for dDD, after removal of random errors due to natural MIC variation, were 87.8%, 88.4%, and 92.2% for the BacT/Alert, Bactec, and VersaTREK systems, respectively. No very major errors (VME) were observed, and major error (ME) rates were 3.0%, 2.3%, and 1.7%, respectively. Incubation of the dDD test samples for 6 h compared to incubation for 16 to 18 h resulted in 19.9% of tests having too light of growth to allow reading of zones of inhibition. Among the evaluable dDD tests, CA values were 58.9%, 76.6%, and 73.2% for the isolates seeded into the BacT/Alert, Bactec, and VersaTREK systems, respectively. VME rates for isolates seeded into these systems were 2.2%, 1.8%, and 3.0%, respectively, and ME rates were 25.4%, 6.1%, and 2.8%, respectively, at the 6-h reading. The best performance of dDD was found for blood cultures with bacterial concentrations in the range of 7.6 × 107 to 5.0 × 108 CFU/ml; CA values ranged from 94.7 to 96.2% for these concentrations after 18 h of incubation and from 76.9 to 84.1% after 6 h of incubation. These preliminary data demonstrate the potential accuracy of dDD testing by the clinical laboratory.
引用
- Chandrasekaran S, Abbott A, Campeau S, et al. Direct-from-Blood-Culture Disk Diffusion To Determine Antimicrobial Susceptibility of Gram-Negative Bacteria: Preliminary Report from the Clinical and Laboratory Standards Institute Methods Development and Standardization Working Group. J Clin Microbiol. 2018. PMID: 29305540. doi:10.1128/JCM.01678-17
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。