液相色谱串联质谱法测定脑脊液胱抑素C
本研究建立并验证了基于液相色谱串联质谱(LC-MS/MS)的脑脊液胱抑素C(CST3)定量方法,评估了精密度、准确度和回收率,并与免疫分析法比较。LC-MS/MS标准曲线线性良好(r²=0.99),批内和批间变异均<10%;LC-MS/MS测得平均CST3浓度显著高于免疫分析法。脑脊液存在25KDa糖基化CST3,去糖基化可减少该组分但不改变免疫法测得浓度。
要点
- LC-MS/MS法测定脑脊液CST3精密度好,批内和批间变异均<10%。
- LC-MS/MS测得CST3平均浓度显著高于免疫分析法。
- 脑脊液含25KDa糖基化CST3,去糖基化不影响免疫法测定值。
核心结论
- LC-MS/MS法测定脑脊液CST3稳健且检测能力更高,可用于神经系统疾病的诊断和监测。
研究设计与样本
- 方法学验证研究:建立LC-MS/MS测定脑脊液CST3,验证精密度、准确度和回收率,并与免疫分析法比较。
- 用Western blotting和凝集素印迹分析脑脊液CST3糖基化。
主要发现
- LC-MS/MS测定标准CST3的线性标准曲线与赋值相关(r²=0.99)。
- 批内和批间变异均<10%。
- LC-MS/MS测得CST3平均浓度显著高于免疫分析法,但两者存在相关性。
- Western blotting显示脑脊液存在25KDa组分和CST3单体(14KDa),去糖基化后25KDa组分减少。
- 凝集素印迹显示唾液酸酶处理的脑脊液中有25KDa糖基化蛋白,去糖基化后减少;去糖基化不改变免疫法测定的CST3浓度。
关键数值与推荐
- LC-MS/MS标准曲线r²=0.99。
- 批内和批间变异均<10%。
临床怎么用
- LC-MS/MS法可用于脑脊液CST3的准确定量,检测能力优于免疫分析法。
- 该方法有望用于神经系统疾病的诊断和监测。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Cystatin C (CST3), a ubiquitously expressed cysteine protease inhibitor, is implicated in several neurological diseases. Here, we have developed an accurate CST3 measurement system based on liquid chromatography tandem mass spectrometry (LC-MS/MS). METHODS: LC-MS/MS based measurement for CSF CST3 was validated by determination of assay precision, accuracy and recovery. The values were compared with those measured by immunoassay. Glycosylation of CST3 in CSF was analyzed by Western blotting and lectin blotting. RESULTS: Measuring standard CST3 by LC-MS/MS produced a linear standard curve that correlated with assigned values (r2=0.99). Both intra- and inter-assay variation was <10%. Although showed a correlation, the average CST3 concentration measured by LC-MS/MS was significantly higher than that of immunoassay. Western blotting showed the presence of a 25KDa species along with CST3 monomer (14KDa) in CSF. The volume of 25KDa species was decreased by deglycosylation. Lectin blotting revealed a 25KDa glycosylated protein in sialidase-treated CSF, which was decreased by deglycosylation. However, deglycosylation did not alter CST3 concentration measured by immunoassay. CONCLUSIONS: Our results suggest that LC-MS/MS-based CST3 measurement is a robust method with higher detection ability. Such method could be useful for the diagnosis and monitoring of neurological diseases.
引用
- Matsuda C, Shiota Y, Sheikh AM, et al. Quantification of CSF cystatin C using liquid chromatography tandem mass spectrometry. Clin Chim Acta. 2017. PMID: 29246666. doi:10.1016/j.cca.2017.12.016
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。