文献速递临床免疫质量管理与危急值

Beckman Coulter UniCel DxI 800 促甲状腺激素(第3代国际标准)检测方法的性能特征

Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay.
PubMed 文献 · Clin Chim Acta · 2017-12-21

本研究对Beckman Coulter基于WHO第3代TSH国际标准重新配方的TSH检测方法进行了全面性能评估,包括精密度、线性、可报告范围、稳定性、灵敏度及常见干扰。结果显示批内和批间CV均<10%,LOQ、LOD和LOB分别为0.0047μIU/ml、0.0012μIU/ml和0.0005μIU/ml,与上一代方法及其他平台结果总体可比,结论认为该检测方法精密、高灵敏且可接受用于临床检测。

要点

  • 批内和批间CV在所有评估浓度下均<10%,精密度良好。
  • LOQ为0.0047μIU/ml(10% CV),LOD为0.0012μIU/ml,LOB为0.0005μIU/ml。
  • 与上一代方法及其他平台比较存在可变偏倚,但总体结果可比。

核心结论

  • UniCel DxI 800的TSH3检测方法精密、高度灵敏,与上一代方法可比,可接受用于临床检测。

研究设计与样本

  • 使用3个浓度的商业质控品评估批内和批间精密度。
  • 使用混合患者标本测定线性、可报告范围、稳定性、灵敏度及对常见干扰的易感性。
  • 在5个不同平台上使用47例患者标本评估批间变异性。

主要发现

  • 批内和批间CV在所有评估浓度下均<10%。
  • LOQ为0.0047μIU/ml(10% CV),LOD为0.0012μIU/ml,LOB为0.0005μIU/ml。
  • 与上一代检测方法及其他平台比较时观察到可变偏倚,但TSH3总体结果可比。

关键数值与推荐

  • 批内和批间CV均<10%。
  • LOQ为0.0047μIU/ml(10% CV),LOD为0.0012μIU/ml,LOB为0.0005μIU/ml。

临床怎么用

  • 该检测方法性能满足临床检测要求,可用于患者TSH测定。
  • 与上一代方法及其他平台比较时需注意可能存在可变偏倚,结果解释时应考虑方法差异。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed. METHODS: Intra-day and inter-day precision was evaluated using 3 concentrations of commercial quality control material. Linearity, reportable range, stability, sensitivity and susceptibility to common inferences were determined using pooled patient specimens. Inter-assay variability was assessed across 5 different platforms (n=47 patient specimens). RESULTS: Intra-day and inter-day CVs were <10% at all concentrations evaluated. The LOQ, LOD and LOB were 0.0047μIU/ml (10% CV), 0.0012μIU/ml and 0.0005μIU/ml, respectively. Variable bias was observed for the TSH3 assay when evaluated against the previous generation assay and other platforms, but overall TSH3 gave comparable results. CONCLUSIONS: The TSH3 assay for UniCel DxI 800, is precise, highly sensitive and comparable to the previous generation assay. The assay is acceptable for clinical testing.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Winston-McPherson GN, Samraj AN, Poster K, et al. Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay. Clin Chim Acta. 2017. PMID: 29275143. doi:10.1016/j.cca.2017.12.029

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。