气动运输系统运送血液样本的安全性的系统综述
共纳入39项研究,其中仅12项针对住院患者(主要为ICU)。由于研究异质性高,现有证据不足以证明PTS运送血液样本的安全性。作者建议实验室测量并记录实际加速度力,建立质量目标阈值,并呼吁制定国际统一建议。
要点
- 现有证据因高度异质性无法证明PTS运送血液样本的安全性。
- 仅少数参数受临床显著影响:外源性血栓弹力图凝血时间、血气pO2、Multiplate分析、溶血指数。
- 实验室应测量并记录PTS实际加速度力,建立质量目标阈值。
核心结论
- 由于研究高度异质性,现有文献无法为PTS运送血液样本的安全性提供证据。
主要发现
- 39项研究中仅12项针对住院患者,主要为ICU患者。
- 血气、血液学和临床生化研究较充分,而凝血、旋转血栓弹力图和急性病患者的血小板功能各仅1项研究。
- 仅少数参数受临床显著影响:外源性血栓弹力图凝血时间、血气pO2、Multiplate分析和溶血指数。
临床怎么用
- 实验室应测量并记录现有PTS的实际加速度力,并建立质量目标阈值(如加速度矢量和)。
- 计算机建模可用于评估未来PTS安装。
局限性
- 纳入研究高度异质性,无法提供安全性证据。
- 针对住院患者(尤其ICU)的研究数量有限。
- 部分分析类型(凝血、旋转血栓弹力图、血小板功能)研究极少。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Pneumatic transportation systems (PTSs) are increasingly used for transportation of blood samples to the core laboratory. Many studies have investigated the impact of these systems on different types of analyses, but to elucidate whether PTSs in general are safe for transportation of blood samples, existing literature on the subject was systematically assessed. METHODS: A systematic literature review was conducted following the preferred reporting items for systematic reviews and metaanalyses (PRISMA) Statement guidelines to gather studies investigating the impact of PTS on analyses in blood samples. Studies were extracted from PubMed and Embase. The search period ended November 2016. RESULTS: A total of 39 studies were retrieved. Of these, only 12 studies were conducted on inpatients, mainly intensive care unit patients. Blood gases, hematology, and clinical chemistry were well investigated, whereas coagulation, rotational thromboelastometry, and platelet function in acutely ill patients were addressed by only 1 study each. Only a few parameters were affected in a clinically significant way (clotting time parameter in extrinsic system thromboelastometry, pO2 in blood gas, multiplate analysis, and the hemolysis index). CONCLUSIONS: Owing to their high degree of heterogeneity, the retrieved studies were unable to supply evidence for the safety of using PTSs for blood sample transportation. In consequence, laboratories need to measure and document the actual acceleration forces in their existing PTS, instituting quality target thresholds for these measurements such as acceleration vector sums. Computer modeling might be applied to the evaluation of future PTS installations. With the increasing use of PTS, a harmonized, international recommendation on this topic is warranted.
引用
- Nybo M, Lund ME, Titlestad K, et al. Blood Sample Transportation by Pneumatic Transportation Systems: A Systematic Literature Review. Clin Chem. 2017. PMID: 29222338. doi:10.1373/clinchem.2017.280479
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。