两种全自动免疫分析系统检测可提取核抗原自身抗体的分析与临床比较
该前瞻性研究纳入疑似系统性自身免疫性风湿病患者,平行使用间接免疫荧光法(IIFA)、化学发光免疫分析法(CIA)和荧光酶免疫分析法(FEIA)检测抗核抗体。结果显示CIA敏感性99%、FEIA敏感性92%、IIFA敏感性90%;特异性分别为76%、84%和76%。两种固相免疫分析法均显著优于IIFA,CIA敏感性更高,FEIA特异性更高。
要点
- CIA检测抗核抗体敏感性99%,高于FEIA的92%和IIFA的90%。
- FEIA特异性84%,高于CIA和IIFA的76%。
- 两种固相免疫分析法总体优于IIFA,差异主要源于截断值。
核心结论
- FEIA和CIA两种固相免疫分析法检测可提取核抗原自身抗体均显著优于IIFA。
- CIA敏感性更高,FEIA特异性更高。
- ROC分析显示两种固相免疫分析法差异的主要贡献因素为截断值。
研究设计与样本
- 前瞻性研究纳入疑似系统性自身免疫性风湿病未选择人群,平行检测IIFA、CIA(QUANTA Flash CTD Screen Plus,BIO-FLASH®)和FEIA(EliA CTD Screen,Phadia® 250)。
- 回顾性验证队列包括ANA相关风湿病患者和健康对照。
主要发现
- 前瞻性队列中IIFA、CIA和FEIA敏感性分别为90%、99%和92%。
- 特异性分别为76%、76%和84%。
- 三种方法总符合率:IIFA与CIA为75.2%,IIFA与FEIA为79.2%,FEIA与CIA为85.4%。
- CIA的AUC为0.95,FEIA为0.93,两者无显著差异。
- 回顾性各ANA相关风湿病队列中,两种CTD筛查方法结果相似。
关键数值与推荐
- CIA敏感性99%,FEIA敏感性92%,IIFA敏感性90%。
- CIA特异性76%,FEIA特异性84%,IIFA特异性76%。
- IIFA与CIA总符合率75.2%,IIFA与FEIA为79.2%,FEIA与CIA为85.4%。
- CIA的AUC为0.95,FEIA的AUC为0.93。
临床怎么用
- 在疑似系统性自身免疫性风湿病患者的抗核抗体筛查中,可考虑用全自动固相免疫分析法替代IIFA。
- 若需更高敏感性可优先选择CIA,若需更高特异性可优先选择FEIA。
- 截断值设置是影响两种固相免疫分析法性能差异的关键因素。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Detection of antinuclear antibodies (ANA) by indirect immunofluorescence assay (IIFA) is increasingly substituted by fully automated solid phase immunoassays. This study evaluated the performance of an automated chemiluminescence immunoassay (CIA) and fluorescence enzyme immunoassay (FEIA) and compared their performance to that of IIFA. METHODS: The study included an unselected prospective study population suspected of systemic autoimmune rheumatic disease. ANA were measured by IIFA, while in parallel sera were tested by CIA QUANTA Flash CTD Screen Plus on the BIO-FLASH® and FEIA EliA CTD Screen on the Phadia® 250 system. As validation, retrospective cohorts of patients with ANA-associated rheumatic disease (AARD) and healthy controls were tested. RESULTS: Prospectively, sensitivity of IIFA, CIA and FEIA was 90%, 99% and 92%, respectively. Specificity was 76%, 76% and 84%, respectively. Total percent agreements between the three methods were 75.2% (IIFA vs. CIA), 79.2% (IIFA vs. FEIA) and 85.4% (FEIA vs. CIA). The AUC values were 0.95 for CIA and 0.93 for FEIA and did not significantly differ. Retrospectively in individual AARD cohorts, similar results were obtained comparing both CTD screens. CONCLUSIONS: Both FEIA and CIA CTD screen significantly outperformed IIFA, with a higher specificity for FEIA and higher sensitivity for CIA. Based on ROC analysis, major contributor to the difference between the two solid phase immunoassays was the cut-off.
引用
- van der Pol P, Bakker-Jonges LE, Kuijpers JHSAM, et al. Analytical and clinical comparison of two fully automated immunoassay systems for the detection of autoantibodies to extractable nuclear antigens. Clin Chim Acta. 2017. PMID: 29170107. doi:10.1016/j.cca.2017.11.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。