文献速递临床免疫

单重特异性IgE检测免疫分析法的比较:ImmunoCAP Phadia 250与Immulite 2000 3gAllergy

Comparison of Singleplex Specific IgE Detection Immunoassays: ImmunoCAP Phadia 250 and Immulite 2000 3gAllergy.
PubMed 文献 · Ann Lab Med · 2018-01-01

本研究比较了ImmunoCAP与Immulite两种单重特异性IgE检测法在209例韩国过敏病患者血清中对吸入性和食物性过敏原的检测性能。共纳入902对配对比较,定性、半定量和定量分析显示两系统性能相似,但虾过敏原相关性较低,临床需注意方法学差异。

要点

  • 两系统对多数过敏原的阳性/阴性一致率在75%至96%之间。
  • 除虾外,所有过敏原的Spearman秩相关系数均超过0.7。
  • 临床解读结果时应考虑两种检测方法的方法学差异。

核心结论

  • ImmunoCAP与Immulite系统性能相似,但存在方法学差异需临床关注。

研究设计与样本

  • 使用209例韩国过敏病患者的血清,比较两种单重sIgE检测法对吸入性和食物性过敏原的检测结果,共纳入902对配对比较。

主要发现

  • 定性比较中,阳性和阴性一致率范围为75%(小麦、虾)至96%(链格孢)。
  • 等级一致性(0-6级)匹配良好,除虾外所有过敏原的Spearman秩相关系数超过0.7。
  • 定量比较中,除虾外所有过敏原的组内相关系数大于0.7。

临床怎么用

  • 临床医生在解读sIgE结果时应考虑ImmunoCAP与Immulite之间的方法学差异。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The identification of clinically meaningful specific immunoglobulin E (sIgE) is important for the diagnosis and management of allergic diseases. Various in vitro sIgE detection methods are available worldwide. Depending on the number of antigens that can be tested simultaneously, there are two representative methods: singleplex and multiplex. Singleplex sIgE detection is mainly provided by Thermo Fisher (ImmunoCAP) and Siemens (Immulite). This study aimed to compare the diagnostic agreement of two singleplex sIgE detection assays. METHODS: Sera from 209 Korean patients with allergic disease were used to compare the ImmunoCAP and Immulite assays with respect to the following allergens: inhalant allergens (Dermatophagoides farinae, cat and dog dander, oak, rye grass, mugwort, Alternaria, German cockroach) and food allergens (hen's egg white, cow's milk, wheat, peanut, soybean, and shrimp). Data from 902 paired comparison tests were included for comparisons. Qualitative, semi-quantitative, and quantitative comparisons were performed using statistical analyses. RESULTS: In qualitative comparisons, the positivity and negativity agreements ranged from 75% (wheat, shrimp) to 96% (Alternaria). Class consistency (classes 0-6) was well matched. Spearman's rank correlation coefficients for all allergens except shrimp were over 0.7. In quantitative comparisons, all allergens excluding shrimp showed >0.7 intra-class correlation coefficients. CONCLUSIONS: The ImmunoCAP and Immulite systems showed similar performances. However, clinicians should consider fundamental methodological differences between the assays.

来源:Ann Lab Med,查看原文。版权归原作者所有。

引用

  1. Park KH, Lee J, Sim DW, et al. Comparison of Singleplex Specific IgE Detection Immunoassays: ImmunoCAP Phadia 250 and Immulite 2000 3gAllergy. Ann Lab Med. 2018. PMID: 29071815. doi:10.3343/alm.2018.38.1.23

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。