贝克曼库尔特Access高敏肌钙蛋白I免疫分析的分析性能评估

Analytical evaluation of the new Beckman Coulter Access high sensitivity cardiac troponin I immunoassay.
PubMed 文献 · Clin Chem Lab Med · 2017-01-01

本研究评估了新型化学发光全自动贝克曼库尔特Access hsTnI高敏免疫分析检测心肌肌钙蛋白I的分析性能,使用锂肝素样本,涵盖空白限、检测限、功能灵敏度、线性、不精密度、175名健康献血者第99百分位上限参考值计算及与两种商用cTnI免疫分析比较。结果显示分析性能优秀,第99百分位URL为15.8 ng/L。

要点

  • Access hsTnI的空白限、检测限和功能灵敏度分别为0.14、0.34和1.35 ng/L。
  • 总不精密度为5.4%-6.1%,线性范围0.95-4195 ng/L内r=1.00。
  • 175名健康献血者中第99百分位URL为15.8 ng/L,98.9%可测得cTnI值。

核心结论

  • 贝克曼库尔特Access hsTnI全自动免疫分析的分析性能优秀。

研究设计与样本

  • 使用锂肝素样本评估空白限、检测限、功能灵敏度、线性、不精密度,并在175名健康献血者(平均年龄36±12岁,47%女性)中计算第99百分位URL,与两种商用cTnI免疫分析比较。

主要发现

  • 空白限、检测限和功能灵敏度分别为0.14、0.34和1.35 ng/L。
  • 批内、批间和总不精密度分别为2.2%-2.9%、4.6%-5.4%和5.4%-6.1%。
  • 在0.95至4195 ng/L范围内线性极佳(r=1.00)。
  • 第99百分位URL为15.8 ng/L,173/175健康受试者(98.9%)可测得cTnI值。
  • 与AccuTnI+3一致性良好(r=0.97;平均偏差-9.3%),与Siemens Dimension Vista cTnI一致性较差(r=0.95;平均偏差-55%)。

关键数值与推荐

  • 空白限0.14 ng/L,检测限0.34 ng/L,功能灵敏度1.35 ng/L。
  • 批内不精密度2.2%-2.9%,批间4.6%-5.4%,总5.4%-6.1%。
  • 线性范围0.95-4195 ng/L,r=1.00。
  • 第99百分位URL为15.8 ng/L。
  • 与AccuTnI+3平均偏差-9.3%,与Siemens Dimension Vista cTnI平均偏差-55%。

临床怎么用

  • 该全自动高敏肌钙蛋白I免疫分析分析性能优秀,可用于临床检测。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: This study was aimed to evaluate the analytical performance of the novel chemiluminescent and fully-automated Beckman Coulter Access hsTnI high-sensitivity immunoassay for measurement of cardiac troponin I (cTnI). METHODS: The study, using lithium heparin samples, included assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, linearity, imprecision (within run, between-run and total), calculation of 99th percentile upper reference limit (URL) in 175 healthy blood donors (mean age, 36±12 years; 47% women) and comparison with two other commercial cTnI immunoassays. RESULTS: The LOB, LOD and functional sensitivity of Access hsTnI were 0.14, 0.34 and 1.35 ng/L, respectively. The within-run, between-run and total imprecision was 2.2%-2.9%, 4.6%-5.4%, and 5.4%-6.1%, respectively. The linearity was excellent in the range of cTnI values between 0.95 and 4195 ng/L (r=1.00). The 99th percentile URL was 15.8 ng/L. Measurable cTnI values were found in 173/175 healthy subjects (98.9%). Good agreement of cTnI values was found with AccuTnI+3 (r=0.97; mean bias, -9.3%), whereas less satisfactory agreement was found with Siemens Dimension Vista cTnI (r=0.95; mean bias, -55%). CONCLUSIONS: The results of our evaluation of the Beckman Coulter Access hsTnI indicate that the analytical performance of this fully-automated immunoassay is excellent.