自动化数字细胞成像分析仪Sysmex DI-60的性能评估
该研究比较了Sysmex DI-60与XN分析仪及手工计数在276例样本(176例异常、100例正常)中的白细胞分类、红细胞分类和血小板估算性能。结果显示DI-60与手工计数白细胞总体一致率为86.0%,验证后一致性极佳,血小板计数与XN高度相关,但DI-60在显著血小板增多时低估血小板。结论认为DI-60性能可接受,有助于优化血液学实验室工作流程。
要点
- DI-60与手工计数白细胞总体一致率86.0%,验证后五分类一致性极佳(加权κ=0.963)。
- DI-60与手工计数中性粒细胞、淋巴细胞、未成熟粒细胞和原始细胞相关性较强。
- DI-60血小板计数与XN高度相关(r=0.945),但显著血小板增多时低估。
核心结论
- DI-60在白细胞分类、红细胞分类和血小板估算方面的性能似乎可接受,即使在异常样本中经审核后也有所改善。
- DI-60有助于优化血液学实验室工作流程并减少手工工作量。
研究设计与样本
- 共纳入276例样本(176例异常和100例正常),比较DI-60与Sysmex XN分析仪和/或手工计数的白细胞分类、红细胞分类和血小板估算结果。
- 红细胞形态学比较依据ICSH分级标准,手工计数按照临床和实验室标准协会指南(H20-A2)进行。
主要发现
- DI-60与手工计数白细胞总体一致率为86.0%。
- DI-60预分类与审核间白细胞五分类一致性极佳(加权κ=0.963)。
- 与手工计数相关性:中性粒细胞r=0.955,淋巴细胞r=0.871,未成熟粒细胞r=0.820,原始细胞r=0.879。
- 基于ICSH分级标准,DI-60与手工计数在红细胞分级上显示显著差异。
- DI-60血小板计数与XN高度相关(r=0.945),但在显著血小板增多样本中低估血小板计数。
关键数值与推荐
- 白细胞总体一致率86.0%;五分类预分类与审核加权κ=0.963。
- 相关性:中性粒细胞r=0.955,淋巴细胞r=0.871,未成熟粒细胞r=0.820,原始细胞r=0.879。
- 血小板计数与XN相关性r=0.945。
临床怎么用
- DI-60可用于血液学实验室白细胞分类、红细胞分类和血小板估算,减少手工工作量。
- 对显著血小板增多样本需注意DI-60可能低估血小板计数。
- 红细胞分级与手工计数存在差异,需结合审核确认。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The Sysmex DI-60 system (DI-60, Sysmex, Kobe, Japan) is a new automated digital cell imaging analyzer. We explored the performance of DI-60 in comparison with Sysmex XN analyzer (XN, Sysmex) and manual count. METHODS: In a total of 276 samples (176 abnormal and 100 normal samples), white blood cell (WBC) differentials, red blood cell (RBC) classification and platelet (PLT) estimation by DI-60 were compared with the results by XN and/or manual count. RBC morphology between pre-classification and verification was compared according to the ICSH grading criteria. The manual count was performed according to the Clinical and Laboratory Standards Institute guidelines (H20-A2). RESULTS: The overall concordance between DI-60 and manual count for WBCs was 86.0%. The agreement between DI-60 pre-classification and verification was excellent (weighted κ=0.963) for WBC five-part differentials. The correlation with manual count was very strong for neutrophils (r=0.955), lymphocytes (r=0.871), immature granulocytes (r=0.820), and blasts (r=0.879). RBC grading showed notable differences between DI-60 and manual counting on the basis of the ICSH grading criteria. Platelet count by DI-60 highly correlated with that by XN (r=0.945). However, DI-60 underestimated platelet counts in samples with marked thrombocytosis. CONCLUSIONS: The performance of DI-60 for WBC differential, RBC classification, and platelet estimation seems to be acceptable even in abnormal samples with improvement after verification. DI-60 would help optimize the workflow in hematology laboratory with reduced manual workload.
引用
- Kim HN, Hur M, Kim H, et al. Performance of automated digital cell imaging analyzer Sysmex DI-60. Clin Chem Lab Med. 2017. PMID: 28672770. doi:10.1515/cclm-2017-0132
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。