Clauss法纤维蛋白原测定:国际标准与商业标准及质控品的互通性评估
该研究评估了国际纤维蛋白原标准品(09/264)、三种商业纤维蛋白原标准品和十种冻干血浆质控品在三种自动凝血分析仪上的互通性,依据CLSI C53A指南比较冻干样本与新鲜冰冻血浆样本的测定偏差。结果显示部分冻干血浆包括国际标准品在STA-Rack Evolution与Sysmex CA1500之间超出95%置信区间,表明其不具互通性。
要点
- 部分冻干血浆包括国际标准品在Clauss法纤维蛋白原测定中不具互通性。
- 三种自动分析仪对新鲜冰冻血浆的纤维蛋白原均值存在显著差异。
- 冻干血浆质控品的互通性问题影响纤维蛋白原测定的溯源链。
核心结论
- 部分冻干血浆包括国际纤维蛋白原标准品在Clauss法纤维蛋白原凝血率测定中不具互通性。
研究设计与样本
- 研究依据CLSI C53A指南,在三种自动分析仪(Sysmex CA1500、STA-Rack Evolution和ACL-Top 700)上使用三种商业凝血酶试剂进行Clauss法凝血率测定,评估国际标准品、三种商业标准品和10种冻干血浆质控品的互通性,并与25份患者新鲜冰冻血浆样本的结果比较。
主要发现
- 冻融对测定结果无影响。
- 新鲜冰冻血浆样本的纤维蛋白原均值在三种自动分析仪间存在显著差异:2.51 g/L(STA-Rack Evolution)、2.25 g/L(ACL-Top 700)和2.20 g/L(Sysmex CA1500)。
- 部分冻干血浆样本包括国际标准品在STA-Rack Evolution与Sysmex CA1500关系间超出95%置信区间。
临床怎么用
- 临床实验室在使用冻干血浆质控品或标准品进行Clauss法纤维蛋白原测定时,需考虑其在不同自动分析仪间的互通性问题。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Many clinical laboratories use a clotting rate assay according to Clauss for the determination of fibrinogen in citrated plasma. The aim of the present study was to assess the commutability of the current International Standard for fibrinogen (coded 09/264), three commercial fibrinogen standards, and 10 freeze-dried plasma quality control samples from various sources. METHODS: Clotting rate assays according to Clauss were performed on three automated instruments (Sysmex CA1500, STA-Rack Evolution and ACL-Top 700), using three commercial thrombin reagents (Siemens, Stago, and Instrumentation Laboratory). Relationships between the results obtained with the three instruments were determined with 25 fresh-frozen plasma samples obtained from patients. The deviations of the assay results obtained with the freeze-dried samples were compared with the deviations obtained with the fresh-frozen samples, according to approved CLSI guideline C53A. RESULTS: Freezing and thawing had no influence on the assay results. There were significant differences in the mean assay results (fibrinogen, g/L) for the fresh-frozen plasma samples between the three automated instruments: 2.51 (STA-Rack Evolution), 2.25 (ACL-Top 700) and 2.20 (Sysmex CA1500). Similar differences were observed for several freeze-dried plasma samples. Some freeze-dried plasma samples, including the International Standard, were out of the 95% confidence interval for the relationship between STA-Rack Evolution and Sysmex CA1500. CONCLUSIONS: Some freeze-dried plasmas including the international standard for fibrinogen are not commutable among automated instruments for fibrinogen clotting rate assays according to Clauss. Our results have consequences for all interested parties in the traceability chain (WHO, industry, external quality assessment schemes, clinical laboratories).
引用
- van den Besselaar AMHP, van Rijn CJJ, Cobbaert CM, et al. Fibrinogen determination according to Clauss: commutability assessment of International and commercial standards and quality control samples. Clin Chem Lab Med. 2017. PMID: 28412717. doi:10.1515/cclm-2016-1088
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。