新型即时检验免疫化学平台实现指南合规的高敏心肌肌钙蛋白T检测
本研究开发了一种基于离心微流控光盘的即时检验(POC)平台,可在30 μL全血样本中11分钟内完成高敏心肌肌钙蛋白T(cTnT)检测。分析性能显示检测下限为3.0 ng/L,在14 ng/L时变异系数(CV)为3.8%,功能灵敏度(CV<10%)为6.8 ng/L,与Elecsys® hs-cTnT检测49份临床血浆样本相关性r=0.9744。结果表明该技术可在POC场景实现高敏cTnT的分析性能。
要点
- 30 μL全血、11分钟 turnaround time 的POC高敏cTnT检测平台。
- 检测下限3.0 ng/L,14 ng/L时CV为3.8%,功能灵敏度6.8 ng/L。
- 与Elecsys® hs-cTnT检测临床血浆样本相关性r=0.9744。
核心结论
- 该离心微流控光盘平台可在即时检验场景实现高敏cTnT检测的分析性能。
研究设计与样本
- 研究为方法学开发与验证,采用离心微流控光盘集成试剂,全血样本经离心分离血浆后与免疫试剂孵育,荧光信号定量cTnT。
主要发现
- 校准曲线覆盖至约8300 ng/L,检测下限为3.0 ng/L。
- 在14 ng/L(Elecsys® hs-cTnT的99百分位值)时CV为3.8%。
- CV<10%的功能灵敏度为6.8 ng/L。
- 与人类骨骼肌肌钙蛋白T无显著交叉反应。
- 49份临床血浆样本与Elecsys® hs-TnT检测相关性r=0.9744。
临床怎么用
- 该平台有望在POC场景提供高敏cTnT检测,支持快速临床决策。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: A multitude of troponin assays for the point-of-care (POC) have been developed showing a lack of analytical sensitivity and precision. We present a new platform solution for the high-sensitivity detection of cardiac troponin T (cTnT) in a 30 μL whole blood sample with a turnaround time of 11 min. METHODS: The immunoassay was completely run in a ready-to-use plastic disposable, a centrifugal microfluidic disc with fully integrated reagents. After the sample application, the assay was automatically processed by separating the cellular blood components via centrifugation, followed by incubation of a defined volume from the generated plasma with the immunoreagents. The fluorescence in the signal zone of a membrane was measured after its washing for the cTnT quantitation. RESULTS: A calibration curve, measured in whole blood samples spiked with native human cTnT, was generated covering a range up to a concentration of approximately 8300 ng/L. The lower detection limit was determined to be 3.0 ng/L. At a concentration of 14 ng/L, the 99th percentile value from the high-sensitivity cardiac troponin T (hs-cTnT) assay in the Elecsys® system, the imprecision (CV) was 3.8%. A CV profile indicated that the functional sensitivity for a CV <10% was 6.8 ng/L. The assay did not show any significant cross-reaction with human skeletal troponin T. We observed an excellent correlation with the hs-TnT Elecsys® assay for 49 clinical plasma samples (r=0.9744). CONCLUSIONS: The described technology shows that an analytical performance for a highly sensitive determination of cTnT can be achieved in a POC setting.
引用
- Lopez-Calle E, Espindola P, Spinke J, et al. A new immunochemistry platform for a guideline-compliant cardiac troponin T assay at the point of care: proof of principle. Clin Chem Lab Med. 2017. PMID: 28375843. doi:10.1515/cclm-2016-1078
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。