光散射技术联合MALDI-TOF质谱快速诊断尿路感染的前瞻性评估

Prospective Evaluation of Light Scatter Technology Paired with Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry for Rapid Diagnosis of Urinary Tract Infections.
PubMed 文献 · J Clin Microbiol · 2017-03-29

该前瞻性研究以培养为参考标准,评估BacterioScan 216Dx光散射设备筛查尿标本,阳性标本直接用于MALDI-TOF MS鉴定和自动化药敏试验。439份尿标本显示筛查灵敏度96.5%、特异度71.4%、阳性预测值45.1%、阴性预测值98.8%;MALDI-TOF MS鉴定灵敏度72.1%、特异度96.9%;药敏总体分类一致率99.2%。

要点

  • 光散射筛查阴性预测值98.8%,阴性结果可短至3小时报告,无需下游培养。
  • 光散射阳性标本直接MALDI-TOF MS鉴定灵敏度72.1%、特异度96.9%。
  • 鉴定成功标本直接药敏分类一致率99.2%,可缩短报告时间超过24小时。

核心结论

  • BacterioScan 216Dx光散射筛查因高灵敏度和阴性预测值,可作为尿培养筛查方法,阴性结果可短至3小时且无需下游处理。
  • 光散射阳性标本联合MALDI-TOF MS鉴定和自动化药敏,在获得鉴定时可缩短报告时间超过24小时。

研究设计与样本

  • 前瞻性评估BacterioScan 216Dx光散射设备,以培养为参考标准,共纳入439份尿标本。
  • 光散射阳性标本用于下游快速MALDI-TOF MS鉴定和自动化系统药敏试验。

主要发现

  • 439份尿标本光散射筛查灵敏度96.5%、特异度71.4%、阳性预测值45.1%、阴性预测值98.8%。
  • 密度选择后悬液MALDI-TOF MS鉴定灵敏度72.1%、特异度96.9%。
  • MALDI-TOF MS鉴定出菌的标本药敏总体分类一致率99.2%。

关键数值与推荐

  • 光散射筛查阴性预测值98.8%,阴性结果可短至3小时报告。
  • MALDI-TOF MS鉴定灵敏度72.1%、特异度96.9%。
  • 药敏总体分类一致率99.2%。

临床怎么用

  • 可用光散射筛查排除阴性尿标本,阴性结果短至3小时报告,减少不必要培养。
  • 光散射阳性标本可直接用于MALDI-TOF MS鉴定和药敏,获得鉴定时可缩短报告时间超过24小时。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Urinary tract infections are one of the most common reasons for health care visits. Diagnosis and optimal treatment often require a urine culture, which takes an average of 1.5 to 2 days from urine collection to results, delaying optimal therapy. Faster, but accurate, alternatives are needed. Light scatter technology has been proposed for several years as a rapid screening tool, whereby negative specimens are excluded from culture. A commercially available light scatter device, BacterioScan 216Dx (BacterioScan, Inc.), has recently been advertised for this application. Paired use of mass spectrometry (MS) for bacterial identification and automated-system-based susceptibility testing straight from the light scatter suspension might provide dramatic improvement in times to a result. The present study prospectively evaluated the BacterioScan device, with culture as the reference standard. Positive light scatter specimens were used for downstream rapid matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) MS organism identification and automated-system-based antimicrobial susceptibility testing. Prospective evaluation of 439 urine samples showed a sensitivity of 96.5%, a specificity of 71.4%, and positive and negative predictive values of 45.1% and 98.8%, respectively. MALDI-TOF MS analysis of the suspension after density-based selection yielded a sensitivity of 72.1% and a specificity of 96.9%. Antimicrobial susceptibility testing of the samples identified by MALDI-TOF MS produced an overall categorical agreement of 99.2%. Given the high sensitivity and negative predictive value of results obtained, BacterioScan 216Dx is a reasonable approach for urine screening and might produce negative results in as few as 3 h, with no downstream workup. Paired rapid identification and susceptibility testing might be useful when MALDI-TOF MS results in an organism identification, and it might decrease the time to a result by more than 24 h.