文献速递临床微生物质量管理与危急值

Gepotidacin 与对照药物对 select 细菌病原体的 MIC 和纸片扩散试验变量分析

Analysis of MIC and Disk Diffusion Testing Variables for Gepotidacin and Comparator Agents against Select Bacterial Pathogens.
PubMed 文献 · J Clin Microbiol · 2017-03-22

本研究采用 CLSI 方法及其变体,测试 gepotidacin 对金黄色葡萄球菌、肺炎链球菌、流感嗜血杆菌和大肠埃希菌的 MIC,以及对金黄色葡萄球菌、肺炎链球菌和化脓性链球菌的纸片扩散法,评估接种量、培养基、温度、CO2 等参数对结果的影响。高接种量和 pH 5.5 影响最大,10% CO2 孵育使所有菌种琼脂稀释 MIC 升高、纸片抑菌圈减小。

要点

  • 高接种量和 pH 5.5 对 gepotidacin 结果影响最大。
  • 10% CO2 孵育升高琼脂稀释 MIC 并减小纸片抑菌圈。
  • pH、琼脂方法、大气条件、温度和血清白蛋白影响较小。

核心结论

  • 高接种量和 pH 5.5 是影响 gepotidacin 体外活性检测的最主要变量。
  • 10% CO2 孵育使 gepotidacin 琼脂稀释 MIC 升高、纸片抑菌圈减小。

研究设计与样本

  • 采用 CLSI 方法及其变体,测试 10 株金黄色葡萄球菌、10 株肺炎链球菌、10 株流感嗜血杆菌和 5 株大肠埃希菌的 MIC。
  • 纸片扩散法测试 30 株金黄色葡萄球菌、15 株肺炎链球菌和 15 株化脓性链球菌。
  • 比较肉汤微量稀释、宏量稀释和琼脂稀释法,并测试培养基、温度、孵育时间、CO2 水平、接种浓度等变量。

主要发现

  • 高接种量和 pH 5.5 条件下,流感嗜血杆菌和肺炎链球菌在肉汤微量稀释中无生长。
  • 10% CO2 孵育时,所有菌种 gepotidacin 琼脂稀释 MIC 升高、纸片抑菌圈直径减小。
  • pH、琼脂方法、大气条件、温度及肉汤法中加血清和白蛋白对结果影响极小。
  • 纸片制造商、琼脂类型及钾和胸苷对肺炎链球菌纸片结果有轻微差异。

临床怎么用

  • 检测 gepotidacin 时应控制接种量和 pH,避免 pH 5.5 条件。
  • CO2 孵育时需注意琼脂稀释 MIC 升高和纸片抑菌圈减小的可能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

This study was conducted to determine the effect of testing parameters on the in vitro activity of gepotidacin, a new triazaacenaphthylene antibacterial agent for the treatment of conventional and biothreat pathogens. CLSI methods, and variations of those methods, were used to test 10 Staphylococcus aureus, 10 Streptococcus pneumoniae, 10 Haemophilus influenzae, and 5 Escherichia coli isolates by MIC and 30 S. aureus, 15 S. pneumoniae, and 15 S. pyogenes isolates by disk diffusion (DD) methods. Levofloxacin and linezolid were tested as comparator agents for MIC and DD methods, respectively. Broth microdilution (BMD), macrodilution (MD), and agar dilution (AD) methods were compared. Variations in media, temperature, incubation time, CO2 level, and inoculum concentration were tested by all methods, and variations in pH, calcium, magnesium, zinc, potassium, thymidine, and polysorbate 80 levels were tested by BMD and DD. The addition of albumin, serum, and lung surfactant was studied by BMD. The variables that impacted the results the most were high inoculum and pH 5.5 (no growth of H. influenzae and S. pneumoniae by BMD). Gepotidacin AD MIC levels were increased and disk zone diameters were decreased for all species in 10% CO2 incubation. The following variables had a minimal effect on gepotidacin results: pH, agar method, atmospheric condition, temperature, and addition of serum and albumin for broth methods. There were also some slight differences in gepotidacin disk results between disk manufacturers and some agar types and also with potassium and thymidine for S. pneumoniae For all other variations, gepotidacin MIC and disk results were considered comparable to reference results.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Koeth LM, DiFranco-Fisher JM, Scangarella-Oman NE, et al. Analysis of MIC and Disk Diffusion Testing Variables for Gepotidacin and Comparator Agents against Select Bacterial Pathogens. J Clin Microbiol. 2017. PMID: 28330893. doi:10.1128/JCM.02366-16

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。