两种全自动免疫分析法检测血清1α,25-二羟基维生素D与LC-MS/MS的比较评估

Evaluation of two fully automated immunoassay based tests for the measurement of 1α,25-dihydroxyvitamin D in human serum and comparison with LC-MS/MS.
PubMed 文献 · Clin Chem Lab Med · 2017-01-01

结果显示DiaSorin总不精密度≤5.2%且与LC-MS/MS强相关、与DEQAS均值一致性好,而IDS iSYS总不精密度达20.1%且高估浓度,尤其在高浓度时。结论认为DiaSorin法因高灵敏度、低不精密度、宽测量范围和良好一致性而有价值。

要点

  • DiaSorin法总不精密度≤5.2%,与LC-MS/MS强相关且与DEQAS均值一致性好。
  • IDS iSYS法总不精密度达20.1%,且高估血清1,25(OH)2 vitD浓度,尤其高浓度时。
  • DiaSorin法因高灵敏度、低不精密度、宽测量范围和良好一致性而有价值。

核心结论

  • DiaSorin全自动免疫分析法检测1,25(OH)2 vitD性能良好,是与LC-MS/MS一致性好的有价值分析选择。
  • IDS iSYS法不精密度高且高估浓度,尤其在高浓度时,性能欠佳。

研究设计与样本

  • 评估DiaSorin和IDS两种全自动免疫分析法检测人血清1,25(OH)2 vitD的性能。
  • 不精密度和线性分别按CLSI EP15-A3和EP6-A指南验证。

主要发现

  • DiaSorin法总不精密度≤5.2%,IDS iSYS法总不精密度达20.1%。
  • DiaSorin法测得的1,25(OH)2 vitD浓度与LC-MS/MS强相关,且与DEQAS样本方法特异性均值一致性好。
  • IDS iSYS法高估人血清1,25(OH)2 vitD浓度,尤其在高浓度时。

关键数值与推荐

  • DiaSorin法总不精密度≤5.2%。
  • IDS iSYS法总不精密度达20.1%。

临床怎么用

  • 对于1,25(OH)2 vitD检测,DiaSorin全自动免疫分析法因高灵敏度、低不精密度、宽测量范围和与LC-MS/MS良好一致性,可作为有价值的分析选择。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: 1α,25-Dihydroxyvitamin D [1,25(OH)2 vitD] is the bioactive form of vitamin D. Due to the very low concentrations of 1,25(OH)2 vitD in the blood and its lipophilic character, measurement of this parameter is analytically challenging. Requiring preceding manual extraction steps before analysis, previous assays have been laborious. METHODS: In the presented study, we evaluated the performance of two immunoassays from DiaSorin and from Immunodiagnostic Systems (IDS) which combine fully automated extraction and measurement of 1,25(OH)2 vitD. Imprecision and linearity were verified according to Clinical and Laboratory Standards Institute EP15-A3 and EP6-A guidelines, respectively. Ninety-three patient serum samples sent to our institute for determination of 1,25(OH)2 vitD, as well as 20 Vitamin D External Quality Assessment Scheme (DEQAS) samples, were used to evaluate correlation and agreement of 1,25(OH)2 vitD measurements between the two immunoassays and with liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS). RESULTS: Total imprecision was 5.2% or less for the DiaSorin test but reached 20.1% for the IDS iSYS test. 1,25(OH)2 vitD concentrations measured with the DiaSorin assay showed a strong correlation with 1,25(OH)2 vitD levels measured by LC-MS/MS and a good agreement with method specific means of DEQAS samples. By contrast, the IDS iSYS test overestimated 1,25(OH)2 vitD concentrations in human serum, particularly at higher concentrations. CONCLUSIONS: Due to its high sensitivity, low imprecision, broad measurement range, and good agreement with 1,25(OH)2 vitD concentrations measured by LC-MS/MS, the DiaSorin test is a valuable analytical option for the determination of 1,25(OH)2 vitD.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。