文献速递临床免疫质量管理与危急值

意大利多中心研究:TSH免疫测定的协调方案

Harmonization protocols for TSH immunoassays: a multicenter study in Italy.
PubMed 文献 · Clin Chem Lab Med · 2017-01-01

该研究纳入7家意大利临床实验室对745份血清样本的TSH测定,并与两种参考实验室用6种常用免疫测定法复测结果比较,同时分析2012-2015年EQA数据。发现常用TSH免疫测定法间存在系统差异,经PCA校准后TSH值变异从中位数13.53%显著降至9.63%。提示EQA方案有助于改善TSH测定协调性。

要点

  • 常用TSH免疫测定法间存在系统差异,可能影响同一患者不同方法结果的解读。
  • 基于PCA的数学校准可显著降低TSH测定值间的变异。
  • EQA方案可用于生成数学公式,帮助临床医生更好比较不同方法的TSH值。

核心结论

  • 常用TSH免疫测定法间存在系统差异。
  • EQA方案有助于改善TSH免疫测定法的协调性,并可产生数学公式供临床医生比较不同方法的TSH值。

研究设计与样本

  • 七家意大利临床实验室测定745份健康受试者和甲状腺疾病患者的血清样本TSH。
  • 样本在-80°C储存2个月后由两家参考实验室用意大利最常用的六种TSH免疫测定法复测。
  • 数据与2012-2015年EQA年度周期中72份质控样本的53,823次TSH测量结果比较。
  • 采用基于主成分分析(PCA)的数学方法对TSH浓度进行重新校准。

主要发现

  • 最常用的市售TSH免疫测定法之间存在系统差异。
  • 临床实验室测定的TSH浓度与参考实验室用相同方法在-80°C储存2个月后测定的结果高度相关。
  • PCA重新校准后TSH值变异从中位数13.53%(10.79%-16.53%)显著降至9.63%(6.90%-13.21%)。

关键数值与推荐

  • EQA质控样本TSH浓度范围约0.1 mIU/L至18.0 mIU/L。
  • PCA校准前TSH值变异中位数为13.53%(10.79%-16.53%),校准后为9.63%(6.90%-13.21%)。

临床怎么用

  • 临床医生可参考EQA方案产生的数学公式,更好地比较不同方法测定的TSH值。
  • 实验室可利用EQA数据改进TSH免疫测定法间的协调性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Systematic difference between thyroid-stimulating hormone (TSH) immunoassays may produce misleading interpretation when samples of the same patients are measured with different methods. The study aims were to evaluate whether systematic differences are present among TSH immunoassays, and whether it is possible to obtain a better harmonization among TSH methods using results obtained in external quality assessment (EQA) schemes. METHODS: Seven Italian clinical laboratories measured TSH in 745 serum samples of healthy subjects and patients with thyroid disorders. These samples were also re-measured by two reference laboratories of the study with the six TSH immunoassays most popular in Italy after 2 months of storage at -80 °C. Moreover, these data were compared to 53,823 TSH measurements, obtained by laboratories participant to 2012-2015 EQA annual cycles in 72 quality control samples (TSH concentrations from about 0.1 mIU/L to 18.0 mIU/L). TSH concentrations were recalibrated using a mathematical approach based on the principal component analysis (PCA). RESULTS: Systematic differences were found between the most popular commercially available TSH immunoassays. TSH concentrations measured by the clinical laboratories were very closely correlated to those measured with the same method by reference laboratories after 2 months of storage at -80 °C. After recalibration using the PCA approach the variation of TSH values significantly decreased from a median pre-calibration value of 13.53% (10.79%-16.53%) to 9.63% (6.90%-13.21%) after recalibration. CONCLUSIONS: Our data suggest that EQA schemes are useful to improve harmonization among TSH immunoassays and also to produce some mathematical formulas, which can be used by clinicians to better compare TSH values measured with different methods.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Clerico A, Ripoli A, Fortunato A, et al. Harmonization protocols for TSH immunoassays: a multicenter study in Italy. Clin Chem Lab Med. 2017. PMID: 28245185. doi:10.1515/cclm-2016-0899

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。