血清催乳素再审视:参数化参考区间与聚乙二醇沉淀法跨平台评估
该研究为六种免疫分析法建立了性别特异性总催乳素和单体催乳素的参数化参考区间,并用27份高催乳素血症和大催乳素血症血清验证。结果表明,当前自动化免疫分析法需基于PEG沉淀的大催乳素筛查来区分真性还是假性血清催乳素升高,两种筛查方法判别能力相似。
要点
- 六种免疫分析法需建立性别特异性总催乳素和单体催乳素参考区间。
- PEG沉淀回收率截断法和单体参考区间法判别能力相似(p=0.089)。
- Cobas和Architect/Kryptor在大催乳素血症血清中误分类数分别最低和最高。
核心结论
- 当前自动化催乳素免疫分析法需基于PEG沉淀的大催乳素筛查以区分真性与假性高催乳素血症。
研究设计与样本
- 使用96份女性与127份男性健康供者血清,为六种免疫分析法确定参数化总催乳素和单体催乳素性别特异性参考区间。
- 用27份高催乳素血症和大催乳素血症血清验证参考区间,并以凝胶过滤色谱法确定单体和大分子催乳素的存在。
主要发现
- 验证血清(高催乳素血症和大催乳素血症,n=27)中,单体参考区间法的分类不一致数均值2.8(95%CI 1.2-4.4),高于PEG后回收率截断法的均值1.8(95%CI 0.8-2.8)。
- 两种单体/大催乳素判别方法差异无统计学意义(p=0.089)。
- 在两种判别方法评估的大催乳素血症血清中,Cobas和Architect/Kryptor催乳素分析法的误分类数分别最低和最高。
临床怎么用
- 临床解读催乳素结果时应考虑大催乳素血症,必要时用PEG沉淀法筛查。
- 单体参考区间法除判别外还可提供易解读的单体催乳素浓度及参考区间。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Hyperprolactinemia diagnosis and treatment is often compromised by the presence of biologically inactive and clinically irrelevant higher-molecular-weight complexes of prolactin, macroprolactin. The objective of this study was to evaluate the performance of two macroprolactin screening regimes across commonly used automated immunoassay platforms. METHODS: Parametric total and monomeric gender-specific reference intervals were determined for six immunoassay methods using female (n=96) and male sera (n=127) from healthy donors. The reference intervals were validated using 27 hyperprolactinemic and macroprolactinemic sera, whose presence of monomeric and macroforms of prolactin were determined using gel filtration chromatography (GFC). RESULTS: Normative data for six prolactin assays included the range of values (2.5th-97.5th percentiles). Validation sera (hyperprolactinemic and macroprolactinemic; n=27) showed higher discordant classification [mean=2.8; 95% confidence interval (CI) 1.2-4.4] for the monomer reference interval method compared to the post-polyethylene glycol (PEG) recovery cutoff method (mean=1.8; 95% CI 0.8-2.8). The two monomer/macroprolactin discrimination methods did not differ significantly (p=0.089). Among macroprolactinemic sera evaluated by both discrimination methods, the Cobas and Architect/Kryptor prolactin assays showed the lowest and the highest number of misclassifications, respectively. CONCLUSIONS: Current automated immunoassays for prolactin testing require macroprolactin screening methods based on PEG precipitation in order to discriminate truly from falsely elevated serum prolactin. While the recovery cutoff and monomeric reference interval macroprolactin screening methods demonstrate similar discriminative ability, the latter method also provides the clinician with an easy interpretable monomeric prolactin concentration along with a monomeric reference interval.
引用
- Overgaard M, Pedersen SM. Serum prolactin revisited: parametric reference intervals and cross platform evaluation of polyethylene glycol precipitation-based methods for discrimination between hyperprolactinemia and macroprolactinemia. Clin Chem Lab Med. 2017. PMID: 28236625. doi:10.1515/cclm-2016-0902