DxI平台新型TSH免疫测定法的多中心评估
本研究在多中心评估Beckman Coulter DxI平台新型TSH免疫测定法的分析性能和临床结果,使用593份血清样本和13份EQA控制品。新方法功能灵敏度达0.0023 mIU/L(20% CV),TSH>0.01 mIU/L时不精密度<5% CV,较旧方法系统性低估平均6.25%。参考人群TSH均值1.694 mIU/L,中位数1.495 mIU/L,97.5百分位3.707 mIU/L。
要点
- 新TSH法功能灵敏度0.0023 mIU/L,TSH>0.01 mIU/L时不精密度<5% CV。
- 新法较旧法系统性低估平均6.25%,低浓度样本差异更大。
- 参考人群TSH均值1.694 mIU/L,97.5百分位3.707 mIU/L。
核心结论
- 新TSH免疫测定法在DxI平台上较旧法有更好分析灵敏度和重现性,并采用WHO第3版IRP 81/563标准及单克隆抗体。
研究设计与样本
- 多中心研究(TSH ELAS Study),使用593份血清样本(来自健康者和甲状腺疾病患者)和13份EQA控制品。
主要发现
- LoB、LoD和20% CV的LoQ分别为0.0004 mIU/L、0.001 mIU/L和0.0023 mIU/L。
- TSH浓度>0.01 mIU/L时不精密度值低于5% CV。
- 新法较旧法系统性低估平均6.25%,主要因低TSH浓度样本方法间差异更大,与新法分析灵敏度更好有关。
- 参考人群(279名健康成人,平均年龄43.6岁,范围20-63岁,138女141男)TSH均值(95%CI)1.694 mIU/L(1.588-1.779),中位数1.495 mIU/L(1.412-1.588),97.5百分位3.707 mIU/L。
关键数值与推荐
- LoB 0.0004 mIU/L,LoD 0.001 mIU/L,20% CV的LoQ 0.0023 mIU/L。
- TSH>0.01 mIU/L时不精密度<5% CV。
- 新法较旧法平均低估6.25%。
- 参考人群TSH均值1.694 mIU/L(95%CI 1.588-1.779),中位数1.495 mIU/L(95%CI 1.412-1.588),97.5百分位3.707 mIU/L。
临床怎么用
- 临床解读TSH结果时需注意新法较旧法系统性偏低约6.25%,尤其低浓度样本。
- 新法采用WHO第3版IRP 81/563标准,实验室应关注标准变更对结果的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM OF THE STUDY: Recently, Beckman Coulter Diagnostics set up a new TSH immunoassay for the automated DxI platform. The aim of this study was to evaluate and compare the analytical performance and clinical results of this method with those of previous method. MATERIAL AND METHODS: A multicenter study (named TSH ELAS Study) was organized using 593 serum samples, collected from healthy subjects and patients with thyroid disorders, and 13 control samples, circulated in an External Quality Assessment (EQA) scheme. RESULTS: The values of LoB and LoD, and LoQ at 20% CV were 0.0004mIU/L, 0.001mIU/L and 0.0023mIU/L, respectively. Moreover, TSH concentrations >0.01mIU/L actually show imprecision values lower than 5% CV. This new TSH assay showed a systematic underestimation (on average of 6.25%) compared to old method, which is mainly due to larger differences between methods for samples with low TSH concentrations, related to the better analytical sensitivity of new compared to old method. In a reference population, including 279 apparently healthy adult subjects, Caucasian volunteers (mean age 43.6years, age 20-63years, 138 women and 141 males) the distribution of TSH concentrations was: mean (CI 95%) 1.694mIU/L (1.588-1.779), median 1.495mIU/L (1.412-1.588mIU/L), 97.5th percentile 3.707mIU/L. CONCLUSIONS: The new TSH immunoassay for DxI platform shows some relevant improvements compared to the previous one: use of the most recent WHO 3rd IRP 81/563 standard and monoclonal antibodies (instead of polyclonal antibodies of the old method), and better analytical sensitivities and reproducibility.
引用
- Dittadi R, Rizzardi S, Masotti S, et al. Multicenter evaluation of the new immunoassay method for TSH measurement using the automated DxI platform. Clin Chim Acta. 2017. PMID: 28189596. doi:10.1016/j.cca.2017.02.006
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。