微孔板法测定经典与旁路补体活性
该研究开发并验证了一种动力学微孔板溶血试验,可在单稀释度人血浆或血清中定量经典和旁路补体活性。以200名健康成人评估校准品活性并建立参考范围,48例患者样本中经典途径与LIA符合率94%、与ELISA符合率87.5%。该方法可同时检测36份双份样本,标准化好、成本低、耗时少。
要点
- 微孔板溶血试验可单稀释度同时测定经典和旁路补体活性。
- 经典途径与LIA符合率94%,与ELISA符合率87.5%。
- 遗传性或获得性补体缺陷患者活性测定<30%。
核心结论
- 该动力学微孔板溶血试验可同时测定经典和旁路补体活性,标准化好、重复性好、成本低且省时。
研究设计与样本
- 研究开发并验证动力学微孔板溶血试验,以96孔板监测致敏绵羊或未包被兔红细胞溶血,校准品活性参考200名健康成人评估。
主要发现
- 经典途径线性及溶血、胆红素血症和脂血症干扰均已评估。
- 日内和日间精密度相对于市售脂质体免疫分析和ELISA表现满意。
- 检出遗传性或获得性补体缺陷患者,活性测定<30%。
- 48例患者样本中经典途径与LIA符合率94%,与ELISA符合率87.5%。
- 该方法灵敏度优于LIA,成本低于LIA和ELISA,且较既往报道的溶血试验省时。
关键数值与推荐
- 参考范围来自200名献血者。
- 每块96孔板每轮可双份检测36份样本。
- 每日绘制校准曲线将50%溶血时间转换为活性百分比。
临床怎么用
- 可用于单稀释度人血浆或血清中经典和旁路补体活性的定量检测。
- 适合需要同时检测多份样本且关注成本和耗时的实验室场景。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: We developed and validated a kinetic microplate hemolytic assay (HA) to quantify classical and alternative complement activity in a single dilution of human plasma or serum. METHODS: The assay is based on monitoring hemolysis of sensitized sheep (or uncoated rabbit) red blood cells by means of a 96-well microplate reader. The activity of the calibrator was evaluated by reference to 200 healthy adults. The conversion of 50% hemolysis time into a percentage of activity was obtained using a calibration curve plotted daily. RESULTS: The linearity of the assay as well as interference (by hemolysis, bilrubinemia and lipemia) was assessed for classical pathway (CP). The within-day and the between-day precision was satisfactory regarding the performance of commercially available liposome immunoassay (LIA) and ELISA. Patients with hereditary or acquired complement deficiencies were detected (activity was measured <30%). We also provided a reference range obtained from 200 blood donors. The agreement of CP evaluated on samples from 48 patients was 94% with LIA and 87.5% with ELISA. The sensitivity of our assay was better than that of LIA, and the cost was lower than either LIA or ELISA. In addition, this assay was less time consuming than previously reported HAs. CONCLUSIONS: This assay allows the simultaneous measurement of 36 samples in duplicate per run of a 96-well plate. The use of a daily calibration curve allows standardization of the method and leads to good reproducibility. The same technique was also adapted for the quantification of alternative pathway (AP) activity.
引用
- Puissant-Lubrano B, Fortenfant F, Winterton P, et al. A microplate assay to measure classical and alternative complement activity. Clin Chem Lab Med. 2017. PMID: 28125402. doi:10.1515/cclm-2016-0553
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。