文献速递临床血液与体液质量管理与危急值

Sysmex XN 血液分析仪在严重白细胞减少样本中的白细胞计数:全血模式与低白细胞模式的分析性能评估

White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode.
PubMed 文献 · Ann Lab Med · 2026-10-16

本研究依据 EFLM 标准评估 Sysmex XN-9100 全血模式与低白细胞模式在 WBC<0.50×10^9/L 样本中的分析性能,精密度除 WBC<0.2×10^9/L 样本外均达标,线性可接受,两模式高度相关(r=0.99)且一致性几乎完美(κ=0.90),差异样本临床解读一致。

要点

  • 两模式精密度均满足 EFLM 标准,但 WBC<0.2×10^9/L 时精密度下降。
  • 两模式线性在 ±4.8% 允许偏差内可接受。
  • 两模式相关性 r=0.99、平均差异 0.00×10^9/L、κ=0.90,差异罕见但仍需谨慎。

核心结论

  • Sysmex XN-9100 全血模式与低白细胞模式在 WBC<0.50×10^9/L 样本中均表现出可靠的分析性能。

研究设计与样本

  • 依据 EFLM 标准(CV% 5.6)用 4 份样本评估两模式精密度。
  • 在 5 个浓度水平评估线性。
  • 对 650 份白细胞减少样本进行两模式比较,并复核 2 份差异样本。

主要发现

  • 两模式精密度均满足 EFLM 标准,但 WBC<0.2×10^9/L 的样本除外。
  • 两模式线性可接受,在 ±4.8% 允许偏差内。
  • 两模式相关性极高(r=0.99),平均差异极小(0.00×10^9/L),一致性几乎完美(κ=0.90)。
  • 两模式差异样本各 1 例,临床解读一致。

关键数值与推荐

  • 精密度评估采用 EFLM 标准 CV% 5.6。
  • 线性评估允许偏差为 ±4.8%。
  • 低白细胞模式自动切换阈值为 WBC<0.50×10^9/L。

临床怎么用

  • 对 WBC<0.50×10^9/L 样本可信任 XN 两模式计数结果,但 WBC<0.2×10^9/L 时需注意精密度下降。
  • 两模式差异罕见,复核差异样本时临床解读一致,可按常规流程处理。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Sysmex XN-9100 (XN) measures WBC counts in whole blood (WB) mode and automatically switches to low WBC (LW) mode for leukopenic samples (WBC <0.50×109/L). However, the analytical performance of LW mode in such samples has not been evaluated according to standard guidelines. We evaluated the analytical performance of the two modes in samples with WBC <0.50×109/L. METHODS: Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed. RESULTS: WBC counting precision met the EFLM criterion for both modes, except in a sample with WBC <0.2×109/L. Linearity was acceptable and within the allowable deviation of ±4.8% for both modes. Very high correlation (r=0.99), minimal mean difference (0.00×109/L), and almost perfect agreement (κ=0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each. CONCLUSIONS: This is the first study to evaluate the analytical performance of WBC counting in WB and LW modes on XN in samples with WBC <0.50×109/L according to guidelines. The two modes demonstrated reliable analytical performance. As precision decreased at WBC <0.2×109/L and discrepancies between the two modes were rare, careful attention is recommended.

来源:Ann Lab Med,查看原文。版权归原作者所有。