卷积神经网络用于ANA与ANCA间接免疫荧光判读的多中心调整与验证
该多中心研究以主中心97例ANA和100例ANCA样本调整EUROPattern系统的初始CNN算法,并在五中心935例ANA和260例ANCA样本中验证。调整后阳性/阴性分类一致性ANA由94.8%升至99.0%、ANCA由84.0%升至96.0%,验证队列ANA与ANCA阳性/阴性一致率分别为98.2%和95.8%,各中心表现一致。
要点
- 调整后CNN与人工判读的阳性/阴性一致率ANA达99.0%、ANCA达96.0%。
- 验证队列ANA与ANCA阳性/阴性一致率分别为98.2%(κ=0.904)和95.8%(κ=0.887)。
- 核型一致率ANA为85.6%、ANCA为94.2%,滴度±1个倍比稀释一致率83.0%~92.8%。
核心结论
- 标准化CNN系统在多中心ANA/ANCA间接免疫荧光判读中表现与人工判读相当,有望提高方法间一致性。
研究设计与样本
- 用主中心97例ANA和100例ANCA样本调整EUROPattern系统的初始CNN算法。
- 在五个中心用调整后CNN系统与人工判读935例ANA和260例ANCA样本,分析方法间一致性。
主要发现
- 调整后阳性/阴性一致率ANA由94.8%升至99.0%,ANCA由84.0%升至96.0%。
- 调整后核型一致率ANA由86.6%升至89.7%,ANCA由84.0%升至96.0%。
- 验证队列ANA与ANCA阳性/阴性一致率分别为98.2%(κ=0.904)和95.8%(κ=0.887),各中心表现一致。
- 验证队列核型一致率ANA为85.6%、ANCA为94.2%,仅非混合样本核均质型差异有统计学意义(P<0.001)。
- 抗体特异性样本中CNN判读更常检出核斑点型和胞质型(P<0.01),SLE、PBC、SS相关疾病核型检出两法相当。
关键数值与推荐
- 单核型ANA、混合核型ANA和ANCA样本滴度±1个倍比稀释一致率分别为92.8%、85.1%和83.0%(二次加权κ分别为0.775、0.648和0.679)。
临床怎么用
- 该CNN系统可用于ANA/ANCA间接免疫荧光判读的辅助,提升与人工判读的一致性并在多中心间保持稳定。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Indirect immunofluorescence (IIF) is the gold standard for autoantibody detection in autoimmune diseases but remains limited by subjective interpretation. We aimed to adjust and validate a standardized convolutional neural network (CNN)-based interpretation system for antinuclear antibody (ANA) and anti-neutrophil cytoplasmic antibody (ANCA) IIF assays across multiple centers. METHODS: The initial CNN algorithm in the EUROPattern (EPA) system was adjusted using 97 ANA and 100 ANCA samples from the lead center. For validation, 935 ANA and 260 ANCA samples from five centers were interpreted by the adjusted CNN-based system and manual interpretation, and inter-method agreement was analyzed. RESULTS: After adjustment, inter-method agreement for positive/negative classification with manual interpretation improved from 94.8% to 99.0% for ANA and from 84.0% to 96.0% for ANCA; pattern agreement improved from 86.6% to 89.7% and from 84.0% to 96.0%, respectively. In the validation cohort, the adjusted system achieved 98.2% (κ = 0.904) and 95.8% (κ = 0.887) positive/negative agreement for ANA and ANCA, with consistent performance across centers. Overall pattern agreement was 85.6% for ANA and 94.2% for ANCA, with statistically significant discordance only for the nuclear homogeneous pattern in non-mixed samples (P < 0.001); in antibody-specific samples, CNN-based interpretation detected nuclear speckled and cytoplasmic patterns more frequently (P < 0.01). Detection of disease-associated patterns in SLE, PBC, and SS was comparable between methods. Titer agreement within ±1 doubling dilution was 92.8%, 85.1%, and 83.0% for single-pattern ANA, mixed-pattern ANA, and ANCA samples (quadratic weighted κ = 0.775, 0.648, and 0.679, respectively). CONCLUSIONS: A standardized CNN-based system performed comparably to manual interpretation for ANA/ANCA IIF interpretation across multiple centers, showing potential to improve inter-method consistency.
引用
- Jia R, Li Y, Hu J, et al. Adjustment and validation of convolutional neural network for indirect immunofluorescence interpretation of antinuclear and anti-neutrophil cytoplasmic antibodies: A multicenter study. Clin Chim Acta. 2026. PMID: 42805451. doi:10.1016/j.cca.2026.122856
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。