文献速递临床免疫质量管理与危急值

降钙素在细针穿刺洗脱液中检测的分析前与分析验证

Pre-analytical and analytical validation of calcitonin measurement in fine needle aspiration cytology washout.
PubMed 文献 · Clin Chem Lab Med · 2026-09-11

该研究评估三种洗脱溶剂(CytoLyt、NaCl、MultiDiluent)对降钙素免疫分析(Siemens Atellica)的分析性能。MultiDiluent满足所有决策标准(批内CV≤5.5%、批间CV≤9%、线性R²=0.997、中位回收率148%、储存稳定),CytoLyt储存不稳定,NaCl仅满足批内CV和线性。结论推荐MultiDiluent用于非血清基质降钙素检测。

要点

  • MultiDiluent是Atellica降钙素免疫分析在FNAC洗脱液中的最优溶剂。
  • CytoLyt精密度和准确度可但储存不稳定,NaCl性能差。
  • 需用患者来源标本进一步验证推荐流程。

核心结论

  • MultiDiluent是Atellica降钙素免疫分析在非血清基质中检测的最佳洗脱溶剂。

研究设计与样本

  • 验证研究使用CytoLyt、NaCl和MultiDiluent三种洗脱溶剂,加入降钙素标准品后评估精密度、准确度和储存稳定性。
  • 降钙素检测采用血清CALCT免疫分析(Siemens, Atellica),决策标准基于CLSI和厂商手册。

主要发现

  • MultiDiluent中降钙素检测批内CV≤5.5%、批间CV≤9%,系列稀释线性R²=0.997,中位回收率148%且变异有限,储存后浓度变化<10%。
  • CytoLyt中检测精密度和准确度可接受,但储存后不稳定。
  • NaCl中仅批内CV和线性满足标准。

关键数值与推荐

  • MultiDiluent批内CV≤5.5%、批间CV≤9%,线性R²=0.997,中位回收率148%,储存变化<10%。

临床怎么用

  • 对于细胞学结论不明确的甲状腺髓样癌患者,FNAC洗脱液降钙素检测可优先选用MultiDiluent作为溶剂。

局限性

  • 需用患者来源材料进一步验证推荐的检测流程。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Current guidelines advise calcitonin measurements in washout from fine needle aspiration cytology (FNAC) upon inconclusive cytological examination of medullary thyroid cancer (MTC). (Pre-)analytical validation studies of calcitonin measurements in FNAC washout are however missing. In this study, we examined (pre-)analytical aspects of calcitonin measurements using three washout solvents, aiming to provide insights for implementation of calcitonin measurements in FNAC washout as a supplementary test for diagnosing and monitoring patients with MTC. METHODS: Validation studies were performed using CytoLyt, NaCl and MultiDiluent as washout solvents. Precision (intra- and inter-assay coefficients of variation (CV)), accuracy (linearity and recovery) and storage stability of each solvent were assessed after spiking with calcitonin standard. Calcitonin measurements were performed using the serum CALCT immunoassay (Siemens, Atellica). Decision criteria were set based on CLSI and manufacturer's manual. RESULTS: Measurements of calcitonin in MultiDiluent met all decision criteria: intra- and inter-assay CVs were ≤5.5 % and ≤9 %, serial dilutions were linear (R2=0.997) and median recovery was 148 % with limited variation. Furthermore, calcitonin concentrations remained stable (<10 % change) after storage. Measurements in CytoLyt were precise and accurate, but unstable upon storage. Measurements in NaCl only met the criteria for intra-assay CV and linearity. CONCLUSIONS: This study showed that MultiDiluent was the optimal solvent to use for calcitonin measurements in a non-serum matrix using the Atellica calcitonin immunoassay. CytoLyt performed suboptimal, while the performance of NaCl was poor. Further research is needed to validate the recommended work-up using patient-derived material.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Mater A, Heijboer AC, Boelen A, et al. Pre-analytical and analytical validation of calcitonin measurement in fine needle aspiration cytology washout. Clin Chem Lab Med. 2026. PMID: 42720008. doi:10.1515/cclm-2026-1057

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。