重新定义绝经后女性hCG正常值:四种化学发光免疫分析系统的97.5百分位参考区间与99百分位阈值

Redefining "Normal" for hCG in postmenopausal women: 97.5th percentile reference intervals and 99th percentile thresholds across four chemiluminescence immunoassay systems.
PubMed 文献 · Clin Chim Acta · 2026-08-17

本研究纳入460名绝经后女性,最终分析355名,用四种化学发光免疫分析系统检测血清hCG,建立97.5百分位参考区间和99百分位阈值。各系统间阈值存在差异,系统B最高、系统C最低,为鉴别垂体来源与病理性hCG升高提供系统特异性定量标准。

要点

  • 四种系统97.5百分位参考区间分别为<7.1、<8.2、<5.7、<5.9 IU/L。
  • 99百分位阈值分别为<8.8、<9.3、<7.1、<7.8 IU/L。
  • hCG与FSH、LH呈中等相关,与年龄仅弱正相关。

核心结论

  • 建立的97.5百分位参考区间和99百分位阈值为鉴别垂体来源与病理性hCG升高提供可靠的系统特异性定量标准。

研究设计与样本

  • 纳入460名绝经后女性,最终分析355名,用四种化学发光免疫分析系统检测血清hCG。
  • 按CLSI EP15-A2评估分析精密度,hCG升高样本经PEG沉淀排除巨复合物干扰。
  • 采用非参数法估计单侧97.5百分位和99百分位上限。

主要发现

  • 四种系统分析精密度均满足分析性能标准。
  • hCG与年龄呈弱正相关(Spearman r=0.146-0.198,均P<0.01)。
  • hCG与FSH呈中等相关(r=0.341,95%CI:0.246-0.430),与LH呈中等相关(r=0.428,95%CI:0.339-0.509)。
  • 系统B的97.5百分位参考区间和99百分位阈值最高,系统C最低。

关键数值与推荐

  • 系统A(Beckman UniCel DxI 800)97.5百分位参考区间<7.1 IU/L,99百分位阈值<8.8 IU/L。
  • 系统B(Siemens Atellica IM 1600)97.5百分位参考区间<8.2 IU/L,99百分位阈值<9.3 IU/L。
  • 系统C(Roche Cobas e 801)97.5百分位参考区间<5.7 IU/L,99百分位阈值<7.1 IU/L。
  • 系统D(Abbott ARCHITECT i2000SR)97.5百分位参考区间<5.9 IU/L,99百分位阈值<7.8 IU/L。

临床怎么用

  • 绝经后女性hCG结果应结合所用免疫分析系统的特异性参考区间和阈值判读。
  • 系统间阈值差异明显,不宜跨系统直接套用统一正常值。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND OBJECTIVE: Physiological human chorionic gonadotropin (hCG) elevations in postmenopausal women are prone to misclassification as pathological. Substantial inter-system variation exists among immunoassays, and system-specific reference intervals (RIs) for this population have not been established. This study aimed to establish 97.5th percentile RIs and 99th percentile thresholds for serum hCG in healthy postmenopausal women across four chemiluminescence immunoassay (CLIA) systems. METHODS: A total of 460 postmenopausal women were enrolled. Serum hCG was measured using four immunoassay systems: Beckman UniCel DxI 800 (A), Siemens Atellica IM 1600 (B), Roche Cobas e 801 (C), and Abbott ARCHITECT i2000SR (D). Analytical precision was evaluated according to Clinical and Laboratory Standards Institute (CLSI) EP15-A2. Samples with hCG > 5.0 IU/L (> 6.0 IU/L for System B) underwent polyethylene glycol (PEG) precipitation to exclude macro-complex interference. Spearman rank correlation was used to assess correlations between hCG and age, follicle-stimulating hormone (FSH), and luteinizing hormone (LH). One-sided upper reference limits (97.5th and 99th percentiles) were estimated using the non-parametric method. RESULTS: 355 subjects were included in the final analysis. The imprecision of all four systems met the analytical performance criterion. Serum hCG showed poor positive correlations with age (Spearman r = 0.146-0.198, all P < 0.01). Fair correlations were observed between hCG and FSH (r = 0.341, 95% CI: 0.246-0.430) and between hCG and LH (r = 0.428, 95% CI: 0.339-0.509). The 97.5th percentile RIs were < 7.1, < 8.2, < 5.7, and < 5.9 IU/L for Systems A-D, respectively; the 99th percentile thresholds were < 8.8, < 9.3, < 7.1, and < 7.8 IU/L. System B consistently yielded the highest thresholds, while System C produced the lowest. CONCLUSION: The established 97.5th percentile RIs and 99th percentile thresholds provide reliable, system-specific quantitative criteria for differentiating pituitary-derived from pathological hCG elevations.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Zhu X, Wang J, Hou L, et al. Redefining "Normal" for hCG in postmenopausal women: 97.5th percentile reference intervals and 99th percentile thresholds across four chemiluminescence immunoassay systems. Clin Chim Acta. 2026. PMID: 42607919. doi:10.1016/j.cca.2026.121282

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。