文献速递质量管理与危急值临床生化临床免疫

基于均值法的患者实时质控在前瞻性实时实施与验证

Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory.
PubMed 文献 · Clin Chem Lab Med · 2026-07-17

本研究在三级医院实验室通过结构化选择流程和前瞻性事件验证,评估基于均值法(AoN)的患者实时质控(PBRTQC)的操作可行性。使用navify Lab Operations的AoN模块对cobas 8000系列检测的患者结果进行监测,筛选出肌酐、钠、氯、钾、钙和游离甲状腺素等分析物,并纳入甲胎蛋白、AST和ALT作比较。14组参数集共产生217,372个AoN点、1,599个报警合并为300个事件,钙、氯、肌酐(门诊)和钠的有意义事件比例较高(占总事件的38-82%),而AFP和钾主要产生非确认性报警。结论认为AoN-PBRTQC可补充常规室内质控,但性能高度依赖分析物特性和机构特定的患者结果分布。

要点

  • AoN-PBRTQC可补充常规室内质控,但需针对分析物和机构优化。
  • 钙、氯、肌酐(门诊)和钠的有意义事件比例较高(38-82%)。
  • AFP和钾主要产生非确认性报警,性能依赖患者结果分布。

核心结论

  • AoN-PBRTQC可补充常规室内质控用于选定分析物,但性能高度依赖分析物特性和机构特定患者结果分布,需持续监测和定期再评估。

研究设计与样本

  • 前瞻性研究,在三级医院实验室使用navify Lab Operations的AoN模块(简单移动平均算法)进行PBRTQC监测。
  • 候选分析物基于cobas 8000系列检测的回顾性患者结果,通过机构特定算法(考虑分布稳定性、生物学变异和操作特征)选择。
  • PBRTQC参数用回顾性训练数据集优化,并在独立数据集验证,前瞻性验证采用基于事件的框架(含患者样本复测和同步室内质控评估)。

主要发现

  • 14组PBRTQC参数集产生217,372个AoN点,触发1,599个报警并合并为300个事件。
  • 钙、氯、肌酐(门诊)和钠的有意义事件占总事件的38-82%。
  • AFP和钾主要产生非确认性报警。

关键数值与推荐

  • 当总体CV与访视类型特异性CV均值的绝对差超过0.5个百分点时,考虑访视类型分层。
  • 选定的分析物包括肌酐、钠、氯、钾、钙和游离甲状腺素,并纳入AFP、AST和ALT作比较。

临床怎么用

  • 对选定分析物,可将AoN-PBRTQC作为常规室内质控的补充工具。
  • 实施时需根据本机构患者结果分布优化参数,并持续监测和定期再评估性能。

局限性

  • 性能高度依赖分析物特性和机构特定的患者结果分布,可能限制普适性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Practical implementation of patient-based real-time quality control (PBRTQC) requires analyte- and institution-specific optimization. This study evaluated the operational feasibility of average-of-normals (AoN)-based PBRTQC through a structured selection process and prospective event-based validation. METHODS: PBRTQC monitoring was performed using the Average of Normals module in navify® Lab Operations (Roche Diagnostics, Mannheim, Germany), which implements a simple moving average (SMA) algorithm. Candidate analytes were selected using retrospective patient results generated on cobas 8,000 series analyzers (Roche Diagnostics, Mannheim, Germany). Analyte selection was guided by a predefined, institution-specific algorithm incorporating distributional stability, biological variability, and operational characteristics. Visit-type stratification was considered when the absolute difference between the overall CV and the mean visit-type-specific CV exceeded 0.5 percentage points. PBRTQC parameters were optimized using a retrospective training dataset and subsequently verified in an independent dataset. Prospective validation was conducted using an event-based framework incorporating patient sample re-measurement and contemporaneous IQC evaluation. RESULTS: Based on the selection algorithm, creatinine, sodium, chloride, potassium, calcium, and free thyroxine were selected, with α-fetoprotein, aspartate aminotransferase, and alanine aminotransferase included for comparison. Across 14 PBRTQC parameter sets, 217,372 AoN points generated 1,599 alarms consolidated into 300 events. Calcium, chloride, creatinine (outpatient), and sodium showed high proportions of meaningful events (38-82 % of total events), whereas AFP and potassium predominantly generated non-confirmatory alarms. CONCLUSIONS: AoN-based PBRTQC can complement conventional IQC for selected analytes. Its performance is strongly dependent on analyte characteristics and institution-specific patient result distributions, underscoring the need for ongoing monitoring and periodic re-evaluation to ensure sustained performance in routine laboratory practice.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Koo SE, Shim S, Park YH, et al. Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory. Clin Chem Lab Med. 2026. PMID: 42461274. doi:10.1515/cclm-2026-0201

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。