文献速递临床生化质量管理与危急值

引入新型校准品并调整校准设置以实现多种载脂蛋白的长期质谱定量

Introduction of new calibrator lots for quantitative protein mass spectrometry and adjustment of calibration settings for long term multiplex apolipoproteins quantification.
PubMed 文献 · Clin Chem Lab Med · 2026-07-10

本研究介绍了用于LC-MS定量九种载脂蛋白的新型天然人血清校准品,并评估了不同曲线拟合方法对长期定量的影响。新校准品除载脂蛋白C-III外浓度范围更高,线性回归中纳入原点作为数据点可提高校准线稳定性,强制过原点并省略截距可减少斜率变异。结论认为强制过原点的线性回归适用于载脂蛋白质谱定量,能保证长期一致的质量。

要点

  • 新校准品涵盖九种载脂蛋白,除apoC-III外浓度范围均高于旧校准品。
  • 线性回归中纳入原点作为数据点可降低斜率变异和截距。
  • 强制过原点并省略截距后,所有载脂蛋白斜率变异小于11.8%CV,两个质控样本不精密度小于9.1%。

核心结论

  • 在多重蛋白质质谱定量中,采用强制过原点的线性回归校准可获得长期一致的质量,适合载脂蛋白检测的临床使用。

研究设计与样本

  • 方法学研究:使用五种旧天然人血清校准品(已成功用于LC-MS载脂蛋白定量超过6年)和五种新校准品,按标准操作规程赋值,评估不同回归方法。

主要发现

  • 新校准品除apoC-III外所有载脂蛋白浓度范围更高。
  • 线性回归中纳入原点作为数据点可提高校准线稳定性,减少斜率变异和截距。
  • 新校准品重新计算并强制过原点应用后,省略线性回归截距,通常降低斜率变异(所有载脂蛋白小于11.8%CV,apo(a)除外;两个质控样本所有载脂蛋白不精密度小于9.1%)。

关键数值与推荐

  • 旧校准品校准采用不约束截距的线性回归。
  • 评估了包含原点作为数据点的回归和强制过原点的回归。
  • 新校准品在强制过原点条件下用旧校准品重新计算。
  • 强制过原点后所有载脂蛋白斜率变异小于11.8%CV(apo(a)除外),两个质控样本所有载脂蛋白不精密度小于9.1%。

临床怎么用

  • 在载脂蛋白LC-MS定量中可考虑采用强制过原点的线性回归校准,以维持长期检测质量。
  • 更换校准品批次时需重新赋值并评估回归模型对斜率变异和不精密度的影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Calibration is a cornerstone of clinical chemistry test results and maintaining an in-house LDT requires periodic renewal of calibrators. Here we present the introduction of a new set of native human serum calibrators, using different curve fitting approaches for long-term quantitation of nine apolipoproteins with traceability to current metrological traceability chains. METHODS: Five "old" native human serum calibrators (OC), have been used successfully for >6 years for quantitation of apolipoproteins by LC-MS. Five new calibrators (NC) were selected and value assigned by standard operating procedures. Calibration with OC used linear regression without constraining the intercept. In addition, regressions including the origin as a datapoint and forcing the origin were evaluated, and NC were recalculated using the OC under origin-forced conditions. RESULTS: The new calibrator set showed higher concentration ranges for all apos, except apoC-III. Including the origin as a datapoint in the linear regression improved stability of calibration lines (with reduction of slope variation and intercepts). Recalculation of the NCs and their subsequent application with origin forced, omitted intercepts of linear regression and in general reduced slope variations (<11.8 %CV, for all apos, except apo(a) with imprecision <9.1 % for all apos for two QC samples. CONCLUSIONS: In a complex system like the mass spectrometric quantitation of multiple proteins, calibration with linear regression using origin forced appears to be a well-suited approach offering guaranteed long-term consistent quality and clinical use of the apolipoprotein test.