文献速递临床微生物质量管理与危急值

两种采血转移装置与标准方法血培养污染率比较

Comparison of blood culture contamination rates with standard practice versus two blood diversion devices at a single institution.
PubMed 文献 · J Clin Microbiol · 2026-06-29

单中心观察性研究在Mayo Clinic Rochester比较Steripath或Kurin初始标本转移装置与标准血培养采集的污染率。共31,215瓶血培养中410瓶污染,Steripath-DTM污染率0.34%低于标准1.43%(P<0.001),但所有装置利用率均未达75%目标且工作人员满意度低。

要点

  • Steripath-DTM污染率0.34%显著低于标准方法1.43%(P<0.001)。
  • Steripath-syringe和Kurin-syringe污染率分别为1.28%和1.76%,未达<0.5%目标。
  • 所有装置利用率均未达75%目标,平均利用率26%–41.5%。

核心结论

  • Steripath-DTM可降低血培养污染率,但所有测试装置均未达到利用率目标且工作人员满意度低。

研究设计与样本

  • 单中心观察性研究,在Mayo Clinic Rochester对超过255名采血员培训后,收集急诊和住院成人共31,215瓶血培养,其中410瓶污染。

主要发现

  • 标准采集基线污染率为1.29%–2.53%。
  • Steripath-DTM污染率0.34% vs 标准1.43%,P<0.001。
  • Steripath-syringe污染率1.28%,Kurin-syringe污染率1.76%,均未达<0.5%主要终点。
  • Steripath-DTM平均利用率26%,Steripath ISDD(合并DTM和syringe)后续研究利用率26.4%,Kurin研究利用率41.5%,均未达>75%目标。
  • 采血员调查显示对两种Steripath配置满意度低,报告多种人体工效学障碍限制其使用。

关键数值与推荐

  • 主要终点为污染率<0.5%,利用率目标为>75%。

临床怎么用

  • Steripath-DTM虽能降低污染率,但因利用率低和工作人员满意度差,未能常规临床实施。

局限性

  • 单中心观察性研究,结果可能受机构操作习惯影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

UNLABELLED: Blood culture contamination can lead to false-positive blood culture results, potentially contributing to increased costs and unneeded antibiotic utilization. Initial specimen diversion devices (ISDD) have been proposed to reduce blood culture contamination rates. A single-center, observational study conducted at Mayo Clinic Rochester compared blood culture contamination rates using Steripath or Kurin ISDDs versus standard blood culture collection protocols. Over 255 phlebotomists were trained in the use of the ISDDs, with 31,215 blood cultures, 410 of which were contaminated, collected from adults in the emergency department and inpatient wards. Baseline contamination rates were 1.29%-2.53% with standard collection methods. The Steripath ISDD with direct-to-media inoculation of blood culture bottles (Steripath-DTM) reduced blood culture contamination compared to standard practice (0.34% vs. 1.43%, respectively, P < 0.001). Follow-up studies with the alternative Steripath configuration with an integrated syringe (Steripath-syringe), and then the Kurin ISDD with syringe collection (Kurin-syringe), did not meet the primary endpoint of <0.5% contamination (Steripath-syringe: 1.28%; Kurin-syringe: 1.76%). All three ISDD configurations failed to meet our goal of >75% utilization, with an average measured utilization of 26% for the Steripath-DTM, 26.4% for the Steripath ISDD (combined DTM and syringe) in a follow-up study, and 41.5% in the Kurin study. Additionally, a survey of phlebotomists revealed low staff satisfaction with both Steripath ISDD configurations. Staff reported multiple ergonomic barriers limiting the Steripath ISDD's utility, which precluded implementation of the Steripath-DTM in routine clinical practice. IMPORTANCE: We conducted a very large, real-world study of two blood diversion devices to determine whether either would meet goals for reduced blood culture contamination and a high rate of utilization by phlebotomists.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Vieni C, Legried M, Mainella J, et al. Comparison of blood culture contamination rates with standard practice versus two blood diversion devices at a single institution. J Clin Microbiol. 2026. PMID: 42370687. doi:10.1128/jcm.00085-26

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。