文献速递临床生化质量管理与危急值

微采样卡在自动化分析仪上检测尿白蛋白肌酐比的分析性能

Analysis of albumin-to-creatinine ratio on automated analysers using a microsampling card (Capitainer DIP70 card) for the collection of urine.
PubMed 文献 · Clin Chem Lab Med · 2026-06-30

本研究评估了一种新型干尿斑(DUS)装置在自动化分析仪上检测尿白蛋白肌酐比(ACR)的分析性能,并与标准液体尿液分析进行比较。

要点

  • DUS装置与液体尿液ACR结果高度一致,R2均大于0.95。
  • DUS支持室温下分散采集,适用于人群筛查和远程监测。
  • 不同平台间ACR差异中位数较小,但Atellica在3-30 mg/mmol区间差异稍大。

核心结论

  • DUS装置为白蛋白和肌酐测量提供了稳健且可重复的液体尿液采样替代方案。

主要发现

  • DUS样本ACR与液体尿液结果高度一致:Roche Cobas Pro R2=0.997,Siemens Atellica R2=0.958,Abbott Alinity R2=0.991。
  • 在ACR区间(<3 mg/mmol、3-30 mg/mmol、>30 mg/mmol)内,Cobas中位差异分别为0.46、0.25、0.62 mg/mmol。
  • Alinity中位差异分别为-0.87、0.27、-0.24 mg/mmol。
  • Atellica中位差异分别为0.47、2.42、1.33 mg/mmol。

临床怎么用

  • DUS装置可用于人群筛查、远程监测和肾脏病研究中的分散式尿液采集。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Albuminuria is a key biomarker of kidney damage, yet preanalytical challenges with conventional liquid urine testing, limiting its utility for decentralized and large-scale screening. This study evaluated the analytical performance of a novel dried urine spot (DUS) device, for measuring urinary albumin-to-creatinine ratio and assessed its agreement with standard liquid urine analysis. METHODS: One hundred routine urine samples were analysed both as liquid and dried specimens prepared with the DUS device. After drying, discs were eluted and tested on three clinical chemistry platforms (Roche Cobas Pro, Abbott Alinity, Siemens Atellica). Albumin-to-creatinine ratio (ACR) was calculated. Statistical analyses included duplicate precision (coefficient of variation), linear regression, and Bland-Altman comparison. A stability study was conducted under multiple temperature conditions. RESULTS: ACR values obtained from DUS samples showed strong concordance with liquid urine results (Roche Cobas Pro: y=1.024×-1.427; R2=0.997; Siemens Atellica y=0.983 + 0.336; R2=0.958 and Abbott Alinity: y=1.062×-0.360; R2=0.991). Across ACR intervals (<3 mg/mmol, 3-30 mg/mmol >30 mg/mmol), Cobas demonstrated small positive median differences (0.46, 0.25, 0.62 mg/mmol), Alinity showed variable differences (-0.87, 0.27, -0.24 mg/mmol), and Atellica yielded consistently positive, slightly larger differences (0.47, 2.42, 1.33 mg/mmol). CONCLUSIONS: The DUS device provides a robust and reproducible alternative to liquid urine sampling for albumin and creatinine measurement. Its ability to enable decentralized, ambient-temperature collection with strong analytical agreement supports its application in population screening, remote monitoring, and kidney disease research.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。