文献速递临床免疫分子诊断

稳定元素标记与电感耦合等离子体质谱定量检测巨细胞病毒IgG抗体的可行性研究

Determination of cytomegalovirus IgG antibodies using stable element labeling and inductively coupled plasma mass spectrometry: Further evidence of the feasibility of antibody detection.
PubMed 文献 · Clin Chim Acta · 2026-04-30

本研究基于稳定元素标记免疫分析与电感耦合等离子体质谱(ICP-MS)联用技术,建立了一种定量检测血清巨细胞病毒(CMV)IgG抗体的方法。方法验证显示良好线性,最高定量限900 U/mL,低、高浓度样本变异系数分别为4.1%和5.5%,回收率90.62%–102.74%,与罗氏电化学发光法比较ROC曲线下面积为0.908。结果表明该系统满足临床实验室检测要求,为质谱在临床抗体检测中的应用提供了新思路。

要点

  • 建立基于铽元素标记和ICP-MS的CMV IgG定量检测系统,方法学性能良好。
  • 最高定量限900 U/mL,低、高浓度CV分别为4.1%和5.5%,日间变异低于10%。
  • 与罗氏ECLIA比较ROC曲线下面积0.908,诊断性能较强。

核心结论

  • 成功建立基于稳定元素标记和ICP-MS的CMV IgG抗体定量检测系统,满足临床实验室检测要求。

研究设计与样本

  • 采用间接法,将重组CMV抗原固定于磁珠,与血清中目标抗体结合后加入铽标记抗人IgG抗体,通过ICP-MS检测铽信号进行定量分析。
  • 依据相关法规和临床实验室标准进行方法验证及临床样本诊断性能评价。

主要发现

  • 方法线性良好,最高定量限为900 U/mL。
  • 低浓度和高浓度样本的变异系数分别为4.1%和5.5%,干扰和日间变异均低于10%。
  • 回收率范围为90.62%至102.74%。
  • 与罗氏电化学发光免疫分析法比较,受试者工作特征曲线下面积为0.908。

关键数值与推荐

  • 最高定量限900 U/mL。
  • 低、高浓度样本CV分别为4.1%和5.5%。
  • 干扰和日间变异低于10%。
  • 回收率90.62%–102.74%。
  • 与罗氏ECLIA比较ROC曲线下面积0.908。

临床怎么用

  • 该技术可实现CMV IgG抗体的定量检测,尤其适用于高抗体水平的准确定量。
  • 质谱法作为抗体检测新平台,具有高灵敏度、无信号干扰和宽检测范围的潜在优势。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Cytomegalovirus (CMV), a β-herpesvirus with broad infectivity, establishes lifelong latency and poses significant risks to immunocompromised individuals, pregnant women, and newborns. Current diagnostic assays detecting serum immunoglobulin G (IgG) are mainly qualitative or semi-quantitative, limiting accurate quantification of high antibody levels. Inductively coupled plasma-mass spectrometry (ICP-MS), based on stable lanthanide rare-earth element labeling, offers high sensitivity, absence of signal interference, and a broad detection range, making it suitable for biomarker detection. This study proposes a technique that combines element-labeled immunoassay and ICP-MS, providing a novel integration for quantitatively detecting CMV IgG. The system was established based on the indirect method. Recombinant CMV antigens were immobilized on magnetic beads and interacted with target antibodies in serum. Following washing, Tb-labeled anti-human IgG antibodies were introduced to complete the immune complex. Quantitative analysis was performed by detecting the Tb signal through ICP-MS. Method validation and evaluation of diagnostic performance in clinical samples followed relevant regulatory and clinical laboratory standards. The test showed excellent linearity, with a maximum quantification limit of 900 U/mL. The coefficients of variation (CV) for low- and high-concentration samples were 4.1% and 5.5%. Interference and day-to-day variations were below 10%. Recoveries ranged from 90.62% to 102.74%. Receiver operating characteristic curve area versus Roche electrochemiluminescence immunoassay (ECLIA) was 0.908, indicating strong diagnostic performance. In summary, we have successfully established a quantitative detection system for CMV IgG antibodies based on stable element labeling and ICP-MS, which meets the requirements for clinical laboratory testing and represents a novel application of mass spectrometry in clinical practice.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Jiang W, Li H, Sun G, et al. Determination of cytomegalovirus IgG antibodies using stable element labeling and inductively coupled plasma mass spectrometry: Further evidence of the feasibility of antibody detection. Clin Chim Acta. 2026. PMID: 42069206. doi:10.1016/j.cca.2026.121051

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。