血清分离管与运输系统对激素和肿瘤标志物检测影响的比较评估

Comparative Evaluation of the Effects of Serum Separator Tubes and Transport Systems on Hormone and Tumor Marker Measurements.
PubMed 文献 · Ann Lab Med · 2026-03-18

研究评估了三种血清分离管与两种运输系统组合对20种激素和肿瘤标志物检测的影响,采集57名志愿者血样。结果显示雌二醇在传送带运输的Vacutainer与Vacuette管间有临床显著差异(-8.21%),胰岛素在气动运输的Vacuette管中有临床重要负偏倚(-6.83%),其他分析物无显著差异。

要点

  • 雌二醇在传送带运输的Vacutainer与Vacuette管间偏倚-8.21%,达临床显著。
  • 胰岛素在气动运输的Vacuette管中负偏倚-6.83%,与溶血相关。
  • Vacuette管最易发生气动运输诱导的溶血,对应最大胰岛素偏倚。

核心结论

  • Ampulab管和AirLab系统性能与对照相当,但雌二醇和胰岛素结果需谨慎。

研究设计与样本

  • 从57名志愿者采集血样,每份均分至六种管型-运输系统组合,用Cobas e801和UniCel DXI 800检测20种分析物。

主要发现

  • 雌二醇在传送带运输的Vacutainer与Vacuette管间平均百分比偏倚-8.21%,为临床显著差异。
  • 胰岛素在气动运输的Vacuette管中平均百分比偏倚-6.83%,为临床重要负偏倚。
  • 其余分析物未显示临床显著或重要差异。

关键数值与推荐

  • 临床显著差异定义为平均百分比偏倚超过EFLM期望偏倚标准,临床重要差异定义为超过最佳偏倚标准。

临床怎么用

  • 使用Ampulab管和AirLab系统时,需注意雌二醇和胰岛素检测可能受管型与运输方式影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Controlling pre-analytical factors ensures the accuracy of hormone and tumor marker assays. We evaluated the combined effects of using different serum separator tubes (i.e., Ampulab [Soyagreentec]), Vacutainer [Becton Dickinson], and Vacuette [Greiner Bio-One]) and transport methods (i.e., the AirLab high-speed pneumatic tube system [PTS; PazKorea] and a conveyor system) on measurement results of 20 analytes. METHODS: Blood was collected from 57 volunteers and equally distributed into six tubes per participant to cover all possible tube type-transport system combinations. Twenty analytes were measured using Cobas e801 (Roche Diagnostics) and UniCel DXI 800 (Beckman Coulter) analyzers. For each analyte, the same analyzer was used across all samples. To evaluate hemolysis in PTS-transported samples, hemolytic index, AST, potassium, and lactate dehydrogenase levels were measured using Cobas c702 (Roche Diagnostics). Clinically significant or important differences were defined as average percentage biases exceeding European Federation of Clinical Chemistry and Laboratory Medicine desirable and optimal bias criteria, respectively. RESULTS: Estradiol showed a clinically significant difference (-8.21%) in conveyor-transported Vacutainer versus Vacuette tubes and clinically important difference between Ampulab and other tubes as separator gel adsorption was different. Insulin exhibited a clinically important negative bias (-6.83%) in PTS-transported Vacuette tubes owing to insulin-degrading enzyme release from hemolyzed erythrocytes. No other analytes exhibited clinically significant or important differences. Vacuette tubes were the most susceptible to PTS-induced hemolysis, corresponding to the largest insulin bias. CONCLUSIONS: Ampulab tubes and the AirLab system performed comparably to their comparators, though caution is warranted for estradiol and insulin results.

来源:Ann Lab Med,查看原文。版权归原作者所有。