文献速递临床免疫质量管理与危急值

依据IFCC LEAP框架验证细针穿刺洗脱液甲状旁腺激素测定

Validation of parathyroid hormone measurement in fine-needle aspiration biopsy washings according to the IFCC LEAP checklist.
PubMed 文献 · Clin Chim Acta · 2026-03-03

该研究在cobas平台用Elecsys PTH法测定细针穿刺洗脱液PTH,按IFCC LEAP框架完成精密度、线性、检测限、携带污染、基质效应、稳定性与稳健性验证,并与组织病理学金标准比对100例。结果显示分析质量良好,但盐水洗脱液存在浓度依赖性基质效应,蛋白补充后明显减轻;与病理总符合率89.0%。

要点

  • FNAB-PTH按IFCC LEAP验证精密度CV≤10%、线性保持、携带污染可忽略。
  • 盐水洗脱液存在浓度依赖基质效应,中位回收率低浓度46.9%、高浓度76.5%。
  • 100例中FNAB-PTH与组织病理总符合率89.0%,支持临床用于确认甲状旁腺来源。

核心结论

  • 按IFCC LEAP框架验证显示FNAB-PTH分析质量良好,尽管存在浓度依赖性基质效应,仍支持其临床用于确认甲状旁腺组织来源和辅助病灶定位。

研究设计与样本

  • 在cobas平台用Elecsys PTH法分析细针穿刺洗脱液,按IFCC LEAP框架评估精密度、线性、检测限与定量限、携带污染、基质效应、稳定性和稳健性。
  • 比较0.9%盐水与蛋白类稀释剂补充后的测定结果,并以组织病理学为参考标准评估100例的临床性能。

主要发现

  • FNAB-PTH精密度优秀(CV≤10%),线性保持,携带污染可忽略,分析灵敏度可接受。
  • 盐水洗脱液存在明显浓度依赖性基质效应,中位回收率从低浓度的46.9%升至髙浓度的76.5%,蛋白补充后该效应明显减轻。
  • FNAB-PTH在常规分析前条件下稳定,但高浓度样本长时间储存时稳定性下降。
  • 100例中FNAB-PTH与组织病理学高度一致,总符合率89.0%。

关键数值与推荐

  • 精密度CV≤10%。
  • 盐水洗脱液中位回收率:低PTH浓度46.9%,高PTH浓度76.5%。
  • 与组织病理学总符合率89.0%(100例)。

临床怎么用

  • FNAB-PTH可用于确认细针穿刺标本来源于甲状旁腺组织并辅助病灶定位,尤其影像学结果不确定时。
  • 解读盐水洗脱液结果时需考虑浓度依赖性基质效应,必要时采用蛋白补充以减轻该效应。
  • 高浓度样本应避免长时间储存,以保证PTH稳定性。

局限性

  • 原文未报告局限性内容。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Measurement of parathyroid hormone in fine-needle aspiration biopsy washings (FNAB-PTH) is widely used in clinical practice to confirm parathyroid tissue origin and support lesion localization, particularly when imaging findings are inconclusive. However, comprehensive analytical validation of FNAB-PTH according to the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Laboratory Evaluation and Analytical Performance (LEAP) framework has not been performed for this non-standard matrix. METHODS: FNAB washings were analyzed using the Elecsys PTH assay on the cobas platform. Analytical performance was evaluated in accordance with the IFCC LEAP framework, including precision, linearity, limits of detection and quantification, carryover, matrix effects, stability, and robustness. Results obtained in 0.9% saline were compared with measurements after supplementation with a protein-based diluent. Clinical performance was assessed by comparison with histopathology, which served as the reference standard. RESULTS: FNAB-PTH demonstrated excellent precision (CV ≤10%), preserved linearity, negligible carryover, and acceptable analytical sensitivity. A pronounced concentration-dependent matrix effect was observed in saline washings, with median recovery increasing from 46.9% at low to 76.5% at high PTH concentrations; this effect was markedly reduced after protein supplementation. PTH in FNAB washings remained stable under routine preanalytical conditions, although reduced stability was observed in high-concentration samples during prolonged storage. In 100 cases, FNAB-PTH showed high agreement with histopathology (overall agreement 89.0%). CONCLUSIONS: Analytical validation of FNAB-PTH according to the IFCC LEAP framework confirms good analytical quality and supports its clinical utility for confirming parathyroid tissue origin and assisting lesion localization, despite the presence of a concentration-dependent matrix effect.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Kozłowska E, Chudziński W, Kumorek A, et al. Validation of parathyroid hormone measurement in fine-needle aspiration biopsy washings according to the IFCC LEAP checklist. Clin Chim Acta. 2026. PMID: 41786261. doi:10.1016/j.cca.2026.120939

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。