MALDI-TOF MS与化学发光免疫法检测IGF-1的一致性研究
该研究纳入660例0至90岁健康人血清,比较MALDI-TOF MS与CLIA检测IGF-1的一致性。全队列中MALDI-TOF MS结果多数落在CLIA试剂盒参考区间内;108例配对检测显示两法强相关(R2=0.982),ICC>0.80,Bland-Altman平均差-0.25 ng/mL,95%一致性界限-24.54至24.04 ng/mL,在临床可接受范围内。
要点
- MALDI-TOF MS与CLIA检测血清IGF-1呈强相关,ICC>0.80。
- Bland-Altman平均差-0.25 ng/mL,95%一致性界限在临床可接受范围。
- 多数年龄组MALDI-TOF MS结果落在CLIA参考区间内。
核心结论
- MALDI-TOF MS检测血清IGF-1与CLIA具有良好一致性,为其临床应用提供依据。
研究设计与样本
- 纳入660例0至90岁健康人血清样本,用MALDI-TOF MS与CLIA检测IGF-1。
- 全队列评估MALDI-TOF MS结果与CLIA试剂盒年龄和性别特异性参考区间的一致性。
- 108例参与者在相同条件下进行配对检测,以评估方法可比性。
- 采用线性回归、组内相关系数(ICC)和Bland-Altman分析评价一致性。
主要发现
- 多数年龄组MALDI-TOF MS测定的IGF-1浓度落在CLIA建立的参考范围内。
- 两法强相关(R2=0.982),回归方程为IGF-1CLIA=0.972×IGF-1MALDI-TOF MS+5.041。
- 儿童相关系数0.994,成人0.985(P<0.001),ICC值均超过0.80。
- Bland-Altman分析平均差-0.25 ng/mL,95%一致性界限-24.54至24.04 ng/mL。
关键数值与推荐
- 两法相关系数R2=0.982,儿童0.994,成人0.985(P<0.001)。
- ICC值均超过0.80,提示高度一致性。
- Bland-Altman平均差-0.25 ng/mL,95%一致性界限-24.54至24.04 ng/mL,在临床可接受范围内。
临床怎么用
- MALDI-TOF MS可作为CLIA检测血清IGF-1的替代方法,但需注意两法差异在临床可接受范围内。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Insulin-like Growth Factor I (IGF-1) is crucial for growth, metabolism, and cell proliferation. Its detection is significant for diagnosing and monitoring various diseases. Chemiluminescence immunoassay (CLIA) is commonly used for IGF-1 detection but is prone to interference and variability. Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) offers higher specificity and sensitivity but is less clinically applied. This study compared the consistency of serum IGF-1 levels detected by MALDI-TOF MS and CLIA. METHODS: A total of 660 serum samples from healthy individuals aged 0 to 90 years were included and analyzed using MALDI-TOF MS and CLIA methods. Initially, the MALDI-TOF MS results for the full cohort were assessed for agreement with age- and sex-specific reference intervals provided by the CLIA kit. Subsequently, a subset of 108 participants underwent paired testing under identical conditions to evaluate method comparability. The consistency between the two methods was evaluated using linear regression, the intraclass correlation coefficient (ICC), and Bland-Altman analysis. RESULTS: In most age groups, the results of IGF-1 concentration measured by MALDI-TOF MS fell within the reference range established by CLIA. A strong correlation (R2 = 0.982) was observed between the two methods, with the regression equation defined as IGF-1CLIA = 0.972 × IGF-1MALDI-TOF MS + 5.041. The correlation coefficients were 0.994 in children and 0.985 in adults (P < 0.001), and ICC values exceeded 0.80, indicating high consistency. Bland-Altman analysis showed a mean difference of -0.25 ng/mL, with 95% limits of agreement ranging from -24.54 to 24.04 ng/mL, which falls within the clinically acceptable range. CONCLUSIONS: MALDI-TOF MS shows good consistency with CLIA in detecting serum IGF-I levels, providing evidence for its clinical application.
引用
- Yan W, Shi R, Zheng W, et al. Consistency study of IGF-1 detection by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) and chemiluminescence immunoassay (CLIA). Clin Chim Acta. 2026. PMID: 41529806. doi:10.1016/j.cca.2026.120835