促甲状腺激素国际参考制品的互换性对计量溯源至关重要

The commutability of thyroid-stimulating hormone international reference preparations plays an important role in metrological traceability.
PubMed 文献 · Clin Chim Acta · 2026-01-13

该研究依据CLSI EP30-A评估促甲状腺激素(TSH)第3版国际参考制品81/565在11种稀释液中的互换性,涉及29份血清样本和8种临床实验室测量程序。结果显示无单一稀释液能在所有检测平台实现互换,使用不可互换基质赋值可产生约25.5%和-23.2%的显著偏倚。TSH参考制品的互换性对临床TSH赋值和多系统计量溯源至关重要。

要点

  • TSH第3版国际参考制品81/565在11种稀释液中互换性不一,无单一稀释液适用于所有检测平台。
  • 使用不可互换基质为血清池赋值时,CI 1000和Cobas 601检测分别记录约25.5%和-23.2%的显著偏倚。
  • 溶解参考制品的基质对确保互换性构成显著挑战,但影响对TSH多系统计量溯源至关重要。

核心结论

  • TSH国际参考制品的互换性对临床实践中TSH赋值至关重要。
  • 溶解参考制品所用基质对确保互换性构成显著挑战,但其影响对TSH多系统计量溯源至关重要。

研究设计与样本

  • 依据CLSI EP30-A指南评估TSH第3版国际参考制品81/565在11种不同稀释液中的互换性,制备2至3种TSH浓度。
  • 评估涉及29份血清样本和8种临床实验室测量程序。
  • 另制备11种不同TSH浓度的人血清池候选标准物质,用系列复溶参考制品赋值,并用另外2种临床实验室测量程序验证稀释液适用性。

主要发现

  • TSH国际参考制品在11种稀释液中表现出可变的互换性,无单一稀释液在所有检测平台实现互换。
  • 使用不可互换基质为血清池赋值时,CI 1000和Cobas 601检测分别记录约25.5%和-23.2%的显著偏倚。
  • 使用10种临床实验室测量程序(含13种方法)为11种不同TSH浓度候选标准物质赋值, harmonized TSH赋值范围为0.191 μIU/mL至44.985 μIU/mL,不确定度在0.039 μIU/mL至3.523 μIU/mL之间变化。

关键数值与推荐

  • TSH第3版国际参考制品81/565在11种稀释液中互换性可变,无单一稀释液在所有检测平台实现互换。
  • 不可互换基质赋值时CI 1000和Cobas 601检测分别记录约25.5%和-23.2%的显著偏倚。
  • 11种候选标准物质的harmonized TSH赋值范围为0.191 μIU/mL至44.985 μIU/mL,不确定度为0.039 μIU/mL至3.523 μIU/mL。

临床怎么用

  • 临床TSH赋值应考虑国际参考制品在不同检测系统中的互换性,避免使用不可互换基质导致显著偏倚。
  • 选择稀释液时需验证其在不同临床实验室测量程序间的互换性,以保障TSH多系统计量溯源。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: This study aimed to assess the commutability of thyroid-stimulating hormone (TSH) international reference preparations (IRP) among various assays and to elucidate their role in achieving metrological traceability across systems. METHODS: We evaluated the commutability of TSH 3rd IRP 81/565, dissolved in eleven different diluents to achieve two to three different TSH concentrations. The assessment involved 29 serum samples across eight clinical laboratory measurement procedures (CLMP) following the CLSI EP30-A guidelines. Furthermore, 11 different TSH concentrations candidate standard materials prepared from human serum pools (HSPs) were prepared and assigned values from a series of reconstituted IRPs to clarify practical significance of commutable reconstituted IRPs. Two more CLMPs were used to assign values for 11 candidate standard materials using TSH IRP to verify applicability of the diluents. RESULTS: The TSH IRP demonstrated variable commutability in the eleven diluents, with no single diluent achieving commutability across all assay platforms. Notably, we recorded significant biases of approximately 25.5% and -23.2% when utilizing uncommutable matrices(the TSH IRP were not commutable between CLMPs in these matrices) for value assignments of HSPs via CI 1000 and Cobas 601 assays, respectively. Lastly, Harmonized TSH assigned values ranged from 0.191 μIU/mL to 44.985 μIU/mL, with uncertainties varying between 0.039 μIU/mL and 3.523 μIU/mL using 10 CLMP including 13 methods for 11 different TSH concentration candidate standard materials. CONCLUSIONS: The commutability of TSH IRP is crucial for assigning TSH values in clinical practice. The matrices employed for dissolving IRPs present significant challenges in securing commutability, but their influence is vital for TSH multisystem metrological traceability.