多重与单重Luminex检测IL-10和TNF-α的潜在偏倚评估

Assessment for potential bias in multiplexed IL-10 and TNF-α from plex count.
PubMed 文献 · Clin Chem Lab Med · 2026-01-19

研究用R&D Systems Luminex HS Cytokine Panel A检测72份血清,比较3-plex与1-plex批次间IL-10和TNF-α的差异。IL-10在1-plex与3-plex间平均差异为4.2 fmol/L(10.7%),小于批内差异;TNF-α批间差异为16.7%,批内为12%。Bland-Altman分析显示低浓度分析物变异明显,但多重与单重检测混用时可比性可能不受实质影响。

要点

  • IL-10在1-plex与3-plex间平均差异4.2 fmol/L(10.7%),小于批内差异。
  • TNF-α批间差异16.7%,批内差异12%。
  • 低浓度分析物变异明显,尤其IL-10,但混用单重与多重检测的可比性可能不受实质影响。

核心结论

  • 多重检测在低分析物浓度下存在变异,尤其IL-10,但与单重检测混用时可比性可能不受实质影响,尤其在低浓度为主的数据集中。

研究设计与样本

  • 使用R&D Systems Luminex HS Cytokine Panel A检测72份血清样本,跨多个批次,包括IL-10和TNF-α各三个3-plex和两个1-plex批次。
  • 采用配对t检验比较1-plex内、3-plex间以及1-plex与3-plex间的细胞因子值差异,并用Bland-Altman图评估绝对和百分比差异。

主要发现

  • IL-10在1-plex与3-plex间平均差异为4.2 fmol/L(10.7%),小于批内差异。
  • TNF-α批间比较差异为16.7%,批内比较差异为12%。
  • IL-10在所有比较中均出现统计学显著差异。
  • Bland-Altman分析显示低分析物浓度下变异明显。

临床怎么用

  • 在低浓度为主的数据集中,多重与单重Luminex检测结果混用可能具有可比性,但需关注低浓度分析物的变异。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Multiplex arrays offer a high-throughput, cost-effective means for quantifying multiple analytes simultaneously, essential for large-scale biomarker research. However, the shared reactive environment in multiplex assays could lead to variations in results compared to single plex assays, potentially impacting outcomes. This study aimed to explore these differences by examining a "3-plex" cytokine panel vs. single plex assays for tumor necrosis factor-alpha (TNF-α) and interleukin-10 (IL-10). METHODS: Using the R&D Systems Luminex HS Cytokine Panel A, 72 serum samples were tested across several batches, including three "3-plex" and two "1-plex" batches each for IL-10 and TNF-α. We evaluated differences in cytokine values using paired t-tests, comparing within 1-plex, between 3-plex, and between 1-plex and 3-plex assays. Bland-Altman plots visually assessed absolute and percentage differences. RESULTS: IL-10 values were similar between 1-plex and 3-plex assays, showing an average difference of 4.2 fmol/L (10.7 %), which was less than the within plex differences. Conversely, TNF-α showed a 16.7 % difference in the between plex comparison, compared to a 12 % difference in the within plex comparisons. Statistically significant differences emerged mainly for IL-10 across all comparisons. Bland-Altman analyses indicated pronounced variability at low analyte concentrations. CONCLUSIONS: While the multiplex assays demonstrated variation at low analyte levels, especially for IL-10, such differences might not substantially affect comparability when mixed with single plex assays, particularly in datasets dominated by low concentrations.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。