因检测干扰导致FSH假性升高的案例分析与处理策略
本文报告一例30岁继发性闭经女性,其卵泡刺激素(FSH)水平意外升高但与正常卵巢储备和自发月经恢复的临床表现不符。通过系列稀释、聚乙二醇(PEG)沉淀、异嗜性抗体阻断和方法比对等评估,发现PEG沉淀后FSH回收率极低(7.09%),且不同平台检测结果不一致,仅Advia Centaur XP(羊源检测抗体)结果在参考范围内,而Abbott Architect i 2000和Roche Elecsys(鼠源检测抗体)结果偏高,提示针对鼠源检测抗体的自身抗体干扰。本案例强调临床与实验室密切协作识别检测干扰的重要性。
要点
- FSH假性升高需与临床表现核对,不符时警惕检测干扰。
- PEG沉淀后FSH回收率7.09%支持大分子FSH或干扰抗体存在。
- 不同平台抗体来源差异可导致FSH检测结果不一致。
核心结论
- FSH假性升高可能由针对鼠源检测抗体的自身抗体干扰所致,需实验室与临床协作识别。
研究设计与样本
- 单病例报告:30岁女性继发性闭经,FSH升高但与临床不符,进行系列稀释、PEG沉淀、异嗜性抗体阻断及方法比对。
主要发现
- 系列稀释呈非线性,PEG沉淀后FSH回收率为7.09%。
- 异嗜性抗体阻断试验无显著影响。
- Advia Centaur XP(羊源抗体)FSH在参考范围内,Abbott Architect i 2000和Roche Elecsys(鼠源抗体)结果不一致。
关键数值与推荐
- PEG沉淀后FSH回收率7.09%支持大分子FSH或干扰抗体。
- 方法比对显示仅Advia Centaur XP结果在参考范围内。
临床怎么用
- 遇FSH结果与临床表现不符时,应联合实验室排查检测干扰,避免不必要干预。
局限性
- 单病例报告,结论需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Laboratory results are important to patient diagnosis and management. Immunoassay interference can lead to misleading results and inappropriate clinical decisions. We present the case of a 30-year-old woman with secondary amenorrhea and unexpectedly elevated follicle-stimulating hormone (FSH) levels that did not align with her clinical findings, including normal ovarian reserve and spontaneous return of menstruation. Suspecting assay interference, a comprehensive evaluation was undertaken, including serial dilution, polyethylene glycol (PEG) precipitation, heterophile antibody blocking, and method comparison studies. Non-linear serial dilution results and markedly low FSH recovery after PEG precipitation (7.09 %) supported the presence of macro-FSH or interfering antibodies. Heterophile blocking tube tests showed no significant effect, making heterophile interference less likely. However, method comparison revealed discordant FSH values across platforms, with only Advia Centaur XP reporting within the reference range. Notably, Advia Centaur used sheep-derived detecting antibodies, unlike Abott Architect i 2000 and Roche Elecsys, which both used mouse-derived antibodies. This led to the hypothesis of an autoantibody against mouse-derived detecting antibodies as the source of interference. This case underscores the importance of close collaboration between clinicians and laboratory professionals to identify assay interference, ensuring accurate diagnosis and avoiding unnecessary interventions.
引用
- Kanat B, Atikeler G, Azarsiz E, et al. Falsely elevated FSH levels due to assay interference: a case-based approach. Clin Chim Acta. 2025. PMID: 40846164. doi:10.1016/j.cca.2025.120563
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。