iStar高敏心肌肌钙蛋白I全血检测的分析与临床评估
本研究评估了iStar高敏心肌肌钙蛋白I(hs-cTnI)检测的分析性能及其与血浆样本的一致性,并建立了性别特异性99百分位参考上限。结果显示该检测灵敏度高、精密度好、线性优异,全血与血浆高度一致,与Abbott ARCHITECT hs-cTnI方法学相关性高,交叉反应可忽略,适用于全血样本快速诊断急性心肌梗死。
要点
- iStar hs-cTnI检测LoB为0.09 ng/L,LoD为0.31 ng/L
- 99百分位参考上限总体16 ng/L,男性18 ng/L,女性14 ng/L
- 全血与血浆高度一致(斜率0.936),与Abbott方法相关性0.983
核心结论
- iStar hs-cTnI检测具有高灵敏度和精密度,适用于全血样本快速诊断急性心肌梗死。
- 全血与血浆高度一致及性别特异性参考上限支持其在急性冠脉综合征管理中的潜在临床应用。
研究设计与样本
- 使用Drawray iStar 500分析仪评估iStar hs-cTnI检测,依据CLSI指南评价LoB、LoD、LoQ、精密度、线性及样本类型一致性。
- 与Abbott ARCHITECT hs-cTnI检测进行方法学比较,并评估与其他肌钙蛋白的交叉反应。
主要发现
- LoB为0.09 ng/L,LoD为0.31 ng/L。
- LoQ在20% CV时为0.79 ng/L,在10% CV时为1.85 ng/L。
- 99百分位URL附近精密度CV为1.4-4.8%。
- 线性优异(r=1.00),全血与血浆一致性高(斜率=0.936)。
- 与Abbott ARCHITECT hs-cTnI相关系数0.983,与骨骼肌肌钙蛋白I、心肌肌钙蛋白C和心肌肌钙蛋白T交叉反应可忽略。
关键数值与推荐
- 健康个体总体99百分位URL为16 ng/L,男性18 ng/L,女性14 ng/L。
临床怎么用
- 可用于全血样本快速诊断急性心肌梗死。
- 性别特异性参考上限有助于急性冠脉综合征管理。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To evaluate the performance of the iStar high-sensitivity cardiac troponin I (hs-cTnI) assay, focusing on its sensitivity, precision, linearity, and consistency with plasma samples, and to establish sex-specific 99th percentile upper reference limits. METHODS: The iStar hs-cTnI assay was assessed using the Drawray iStar 500 analyzer. Key performance metrics such as the limits of blank (LoB), detection (LoD) and quantitation (LoQ), precision, linearity, and agreement between sample types were evaluated according to Clinical and Laboratory Standards Institute (CLSI) guidelines. A methodological comparison was performed with the Abbott ARCHITECT hs-cTnI assay, and cross-reactivity with others troponins was assessed. RESULTS: The iStar hs-cTnI assay demonstrated robust sensitivity with a LoB of 0.09 ng/L and LoD of 0.31 ng/L. The LoQ was 0.79 ng/L for 20 % coefficient of variation (CV) and 1.85 ng/L for 10 % CV. Precision testing revealed CVs of 1.4-4.8 % near the 99th percentile upper reference limit (URL). The assay exhibited excellent linearity (r=1.00) and high agreement between whole blood and plasma samples (slope=0.936). Methodological comparison with the Abbott ARCHITECT hs-cTnI showed a high correlation coefficient of 0.983. Cross-reactivity with skeletal muscle troponin I, cardiac troponin C, and cardiac troponin T was negligible. In healthy individuals, the overall 99th percentile URL was 16 ng/L, with sex-specific values of 18 ng/L for males and 14 ng/L for females. CONCLUSIONS: The iStar hs-cTnI assay demonstrates high sensitivity and precision, supporting it suitable for the rapid diagnosis of acute myocardial infarction using whole blood samples. Its high agreement with plasma and established sex-specific URL support its potential for clinical use in acute coronary syndrome management.
引用
- Yu JX, Cui K, Li Y, et al. Analytical and clinical evaluation of an automated high-sensitivity cardiac troponin I assay for whole blood. Clin Chem Lab Med. 2025. PMID: 40811682. doi:10.1515/cclm-2025-0143
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。