人血浆醛固酮免疫法高估原因探究:非合成无醛固酮基质研究
研究使用二氯甲烷(DCM)萃取和非合成无醛固酮血浆基质(ALFP),探究直接免疫法高估醛固酮的机制。所有样本直接DiaSorin免疫法结果均高于LC-MS/MS;DCM萃取后以DED复溶结果与LC-MS/MS一致,而以ALFP复溶则高估约50 pmol/L。提示患者样本中内源性蛋白可能结合示踪剂干扰免疫法。
要点
- 直接DiaSorin免疫法测定醛固酮在所有样本中均高于LC-MS/MS。
- DCM萃取后以DED复溶,醛固酮结果与LC-MS/MS匹配良好。
- 以ALFP复溶导致所有样本醛固酮高估约50 pmol/L。
核心结论
- 直接DiaSorin醛固酮免疫法高估的最可能原因是患者样本中常见的内源性蛋白,该蛋白可能通过结合示踪剂干扰免疫测定。
研究设计与样本
- 使用来自eGFR正常患者的血浆样本,覆盖宽醛固酮浓度范围,比较直接DiaSorin免疫法、DCM萃取后LC-MS/MS及不同基质复溶的结果。
- 复溶基质为DiaSorin Liaison内分泌稀释液(DED,有效成分为人血清白蛋白基质)或非合成人无醛固酮血浆(ALFP)。
主要发现
- 直接免疫法测定的醛固酮浓度在所有样本中均高于LC-MS/MS。
- DCM萃取后以DED复溶,醛固酮结果与LC-MS/MS结果匹配良好。
- DCM萃取后以ALFP复溶而非DED,导致所有样本醛固酮高估约50 pmol/L。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Aldosterone is over-estimated by direct immunoassay when compared to assay by LC-MS/MS. We have investigated the mechanism for this over-estimation, using dichloromethane (DCM) extraction and a unique matrix of non-synthetic aldosterone-free plasma (ALFP). METHODS: Samples used were from patients with normal estimated Glomerular Filtration Rate (eGFR). All samples were measured by direct DiaSorin immunoassay, followed by extraction using dichloromethane (DCM) and LC-MS/MS. Post-DCM samples were reconstituted in DiaSorin Liaison Endocrinology Diluent (DED), which is effectively a human serum albumin matrix, or in a unique matrix of non-synthetic human aldosterone-free plasma (ALFP). Patient plasma samples were selected to cover a wide range of aldosterone concentrations. RESULTS: Aldosterone direct immunoassay over-estimated aldosterone concentration in all samples, compared to LC-MS/MS. Post-DCM extraction and reconstitution with DED, aldosterone results matched well with LC-MS/MS results. Reconstitution with our unique ALFP plasma matrix rather than the DED, resulted in an over-estimation of aldosterone by approximately 50 pmol/L in all samples. CONCLUSIONS: In this investigation, we have demonstrated that the most likely source of interference in the direct DiaSorin aldosterone immunoassay is an endogenous protein commonly found in patient samples. This protein probably interferes with the immunoassay by binding to the tracer.
引用
- Stoner S, Mogambi S, Ko L, et al. Investigation of the possible cause of over-estimation of human aldosterone in plasma, using a unique, non-synthetic human aldosterone-free matrix. Clin Chem Lab Med. 2025. PMID: 40742120. doi:10.1515/cclm-2024-1468