文献速递临床微生物质量管理与危急值

VITEK MS PRIME与Bruker Biotyper MALDI-TOF MS系统的性能和工作流程比较

Performance and workflow comparison of the VITEK MS PRIME and Bruker Biotyper MALDI-TOF MS systems.
PubMed 文献 · J Clin Microbiol · 2025-06-20

本研究比较了VITEK MS PRIME(PRIME)与Bruker MALDI Biotyper CA系统(Biotyper)对154株细菌和酵母临床分离株的鉴定性能及工作流程时间,并分析了阳性血培养分离株在短孵育(6-8小时)和常规孵育(18-24小时)下的鉴定表现。结果显示PRIME与Biotyper鉴定性能相当,PRIME在工作流程上可节省手工操作时间,尤其在高通量场景下具有优势。

要点

  • PRIME对挑战分离株的属水平鉴定率为96%(PICKME)和95%(Loop),Biotyper为99%。
  • 短孵育阳性血培养分离株的种水平鉴定率:Biotyper 84%,PRIME PICKME 80%,PRIME Loop 81%。
  • PRIME方法在多靶标分析时手工操作时间更短(Biotyper 53分钟,PRIME PICKME 39分钟,PRIME Loop 40分钟)。

核心结论

  • PRIME与Biotyper系统鉴定性能相当,PRIME工作流程可节省时间,尤其适合高通量场景。

研究设计与样本

  • 研究比较了154株来自不同标本类型的细菌和酵母临床分离株,使用三种方法:Biotyper靶板加牙签、PRIME靶板加PICKME笔尖、PRIME靶板加接种环。
  • 阳性血培养分离株在短孵育(6-8小时)和常规孵育(18-24小时)后,由高经验用户(HEU,n=300)和低经验用户(LEU,n=50)进行分析。
  • 评估了从样本处理到微生物鉴定的工作流程时间。

主要发现

  • PRIME对挑战分离株的属水平鉴定率为96%(PICKME)和95%(Loop),Biotyper为99%。
  • 短孵育阳性血培养分离株的种水平鉴定率:Biotyper 84%,PRIME PICKME 80%,PRIME Loop 81%,但短孵育比常规孵育需要更多重复。
  • 短孵育时,任何方法在用户间种水平鉴定率无差异(HEU 89%-91% vs LEU 79%-85%)。
  • 单靶标流程时间所有方法相当(55-59分钟),而PRIME方法在多靶标分析时手工操作时间更短(Biotyper 53分钟,PRIME PICKME 39分钟,PRIME Loop 40分钟)。

临床怎么用

  • PRIME可用于临床微生物实验室的细菌和酵母快速鉴定,性能与Biotyper相当。
  • PRIME工作流程可减少手工操作时间,尤其在高通量实验室中可提高效率。
  • 短孵育阳性血培养可用于快速鉴定,但可能需要更多重复。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) is widely used for rapid identification of bacteria and yeast in clinical labs. The VITEK MS PRIME ("PRIME") is the latest MALDI-TOF MS system from bioMérieux. This study evaluated PRIME's performance and workflow timing against the MALDI Biotyper CA System by Bruker ("Biotyper"). We compared a collection of 154 bacteria and yeast clinical isolates from various specimen types using three methods: Biotyper target with a toothpick, PRIME target with the PICKME nib, and PRIME target with a loop. Positive blood culture isolates were also analyzed using these methods, with high-experience (HEU, n = 300) and low-experience users (LEU, n = 50) after short (6-8 h) and routine (18-24 h) incubation on agar plates. Workflow timing, from sample processing to organism identification, was assessed to identify time savings. PRIME identified 96% (PICKME) and 95% (Loop) of challenge isolates to genus level, compared with Biotyper at 99%. Short incubation of positive blood culture isolates demonstrated similar species-level identification rates across all methods (Biotyper: 84%, PRIME PICKME: 80%, PRIME Loop: 81%), although more repeats were needed compared with routine incubation. No difference in species identification occurred between users for any method at short incubation (89%-91%, HEU, vs 79%-85% LEU). Single-target process times were comparable for all methods (55-59 min), whereas PRIME methods had shorter hands-on times for analysis of multiple targets (Biotyper: 53 min, PRIME PICKME: 39 min, PRIME Loop: 40 min). These findings highlight the comparable performance of the PRIME and Biotyper systems while demonstrating the potential for time savings with PRIME workflows, particularly in high-throughput settings.IMPORTANCEThis study provides a critical evaluation of the new VITEK MS PRIME MALDI-TOF MS system, comparing its performance and workflow efficiency against the Bruker MALDI Biotyper. The study investigates the success rate of isolate identification from short incubation of positive blood cultures, illustrating the utility of this technique for downstream workflows such as faster reporting of results for patient management and isolate identification for interpretation of rapid phenotypic AST. Analysis of different workflows demonstrated areas for potential time savings, particularly in high-throughput settings. These findings highlight the importance of optimizing MALDI-TOF MS workflows in the era of workforce shortages and lab centralization to enhance rapid pathogen identification and improve patient care.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Bosserman RE, Tarlton NJ, Alvarado K, et al. Performance and workflow comparison of the VITEK MS PRIME and Bruker Biotyper MALDI-TOF MS systems. J Clin Microbiol. 2025. PMID: 40539832. doi:10.1128/jcm.00211-25

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。